REMOTE CONTROL
59
Return parameter <NR1>
Returns bleeder resistor status.
SYSTem:CONFigure:BTRip[:IMMediate]
Set
Description
Trips the power switch trip (circuit breaker) to
turn the unit off (shut down the power).
Syntax
SYSTem:CONFigure:BTRip[:IMMediate]
SYSTem:CONFigure:BTRip:PROTection
Set
Query
Description
Enables/Disables the power switch trip (circuit
breaker) when the OVP or OCP protection settings
are tripped. This setting only applies after power
has been reset.
Syntax
SYSTem:CONFigure:BTRip:PROTection {OFF|ON|0|1}
Query Syntax
SYSTem:CONFigure:BTRip:PROTection?
Parameter
0
<NR1> Disables the power switch trip for
OVP or OCP.
OFF
Disables the power switch trip for OVP or
OCP.
1
<NR1> Enables the power switch trip for
OVP or OCP.
ON
Enables the power switch trip for OVP or
OCP.
Return parameter <Boolean>
Returns power switch trip setting.
SYSTem:CONFigure:CURRent:CONTrol
Set
Query
Description
Sets or queries the CC control mode (local control
(panel), external voltage control, external
resistance control). This setting is applied only
after the unit is reset.
Syntax
Query Syntax
SYSTem:CONFigure:CURRent:CONTrol { 0 | 1 | 2 | 3 }
SYSTem:CONFigure:CURRent:CONTrol?
Parameter/Return <NR1>
Description
0
Local (Panel) control