OPERATION
133
Select CR Dynamic Mode
Background
Constant Resistance mode can be set to dynamic
or static mode. Dynamic mode is used to
automatically set varying load rates.
Ensure the menu is in CR Mode. See page 131.
Configure
Configure
Mode
CC
Dynamic
Range
High
Panel Operation
1.
Press F3 until Dynamic Range mode
is selected.
Dynamic
Static
F1
F5
~
F3
Note
Changing from static to dynamic mode will only affect
the current (active) channel.
Editing CR Dynamic Parameters
CR levels
Dynamic Constant Resistance Mode has two
operating resistance levels, slew rates and timers.
Slew rates determine the speed at which the load
will change from one level to the next.
The timers determine how long the load
module/channel will stay at level 1 or level 2. See
the CR operating description for details about
slew rates and timers, page 63
Ensure the menu is in CR Dynamic Mode. See
page 133.
Configure
Configure
Mode
CC
Dynamic
Range