GPD-X303S Series User Manual
Command Syntax
Command format
ISET<X>:<NR2>NL
1
2 3
4
5
1: command header
2: output channel
3: separator
4: parameter
5: terminator(line feed)
Output channel
1 (CH1) or 2 (CH2)
Parameter Type
Description Example
<Boolean>
boolean logic
0 (off), 1 (on)
<NR1>
integers
0, 1, 2, 3
<NR2>
decimal numbers 0.1, 3.14, 8.5
Terminator
Each command must end with a terminal character
(new line code, ASCII: 0x0A) and each query must
end with a carriage return, ASCII: 0x0D.
Note
Commands are not case-sensitive.
Error Messages
The following error messages might appear when the GPD-X303S
cannot accept the command.
Message contents
Descriptions
a Program
mnemonic too
long
The command length must be 15 characters
or less.
b Invalid character
Invalid characters, such as symbols, are
entered. Example: VOUT#
c Missing
parameter
The parameter is missing from the command.
Example: VSET: (should have a number)
d Data out of range
The entered value exceeds the specification.
Example: VSET:33 (should be
≤
32V)
46
All manuals and user guides at all-guides.com