REMOTE CONTROL
187
[SENSe:]VOLTage:DC:RANGe?
Returns the DC Voltage measurement range.
Parameter: [None] | [MIN | MAX]
[SENSe:]VOLTage:AC:RANGe
Sets the AC Voltage measurement range.
Parameter: (<NRf> | MIN | MAX)
Example: SENS:VOLT:AC:RANG MIN
Set the AC voltage range to lowest range allowed.
[SENSe:]VOLTage:AC:RANGe?
Returns the AC Voltage measurement range.
Parameter: [None] | [MIN | MAX]
[SENSe:]CURRent:DC:RANGe
Sets the DC Current measurement range.
Parameter: Range(<NRf> | MIN | MAX)
Example: SENS:CURR:DC:RANG 10 e-2
Sets the DC current range to 100mA.
[SENSe:]CURRent:DC:RANGe?
Returns the DC Current measurement range.
Parameter: [None] | [MIN | MAX]
[SENSe:]CURRent:AC:RANGe
Sets the AC Current measurement range.
Parameter: Range(<NRf> | MIN | MAX)
Example: SENS:CURR:AC:RANG 10 e-2
Sets the AC current range to 100mA.
[SENSe:]CURRent:AC:RANGe?
Returns the AC Current measurement range.
Parameter: [None] | [MIN | MAX]
[SENSe:]RESistance:RANGe
Sets the 2W resistance measurement range.
Parameter: Range(<NRf> | MIN | MAX)
Example: SENS:RES:RANG 1000
Sets the resistance range to 1k
Ω
.
[SENSe:]RESistance:RANGe?
Returns the 2W resistance measurement range.
Parameter: [None] | [MIN | MAX]