WITE32 Release Notes Version 3.11
CHAPTER 3
FIXED BUGS
The following bugs were discovered in WITE32 version 3.10 or earlier, and fixed in WITE32
version 3.11. The description below explains the bug behavior as it appears in WITE32 version 3.10.
3.1 V2002
Spinstand
1. The
XY Alignment
test may fail with the
"Servo amplifier is not calibrated - signal is too low"
error message.
2. The
XY Alignment
and the
Head Alignment
tests fail in the case of low frequency heads (the heads with data rates
below 160 Mbit/s).
3. The
XY Alignment
test resets the loading radius to the default value in the
Product Configuration
dialog box.
4. The
Y-limit Screw Adjustment
test fails (shows unreachable destination position) if non-zero Head Base
Parameters are specified in the
Product Configuration
dialog box.
5.
The
Product Parameters
specified in V2002.DDT in Inches are different from parameters shown in
Edit
Product
Parameters
dialog by 0.002%. As a result the parameters loaded into spinstand are different from parameters
specified in DDT file by 0.002%.
6.
Impossible to use the
Special Disk Type
setting in the
Product Parameters
dialog box of the
V2002 Spinstand
Alignment Program
, because the
Minimum Radius
and
Maximum Radius
values are not stored in the product
database.
7. After
performing
the
Y-limit Adjustment
there is no software validation of Y-stage positioning and the Y-limit
screw can be hit. In the new version of WITE32, the software always checks if the destination position is
reachable before the Y-stage starts moving such that the Y-stage does not hit the screw and the warning message
is shown when the specified destination position is out of range.
8. The value of the Reset Radius in the Product Parameters dialog box is different after performing
XY Alignment
with zero and non-zero head geometry parameters specified in the Product Parameters dialog box.
9. The
V2002 Balancing
procedure for the
Custom 6 Screws
and
Custom 3 Screws
choices from the
Type
dropdown
list of the
Balancing Parameters
frame instructs to adjust screws for counter-clockwise balancing screw
numeration instead of clockwise balancing screw numeration.
10. Cartridge, which is incorrectly installed on V2002 spinstand could hit the media or spindle. The sequence of
starting the device is modified such that the spinstand checks that the required level of vacuum is provided to the
cartridge before the X and Y motors start moving the V2002 head loader to the loading location. If the vacuum
does not reach the required level due to improperly installed cartridge, the spinstand will not move and will
provide an error message.
11. The V2002 Balancing while executed for rotation speeds higher then 10000 RPM consists of 4 stages in the case
of high imbalance:
•
Screw Calibration for reduced RPM,
•
Balancing for reduced RPM,
•
Screw Calibration for full RPM,
•
Balancing for full RPM.
In WITE32 version 3.10 the Screw Calibration for full RPM shows the wrong instruction to
"Switch to Balancing
Tab to perform Balancing procedure"
instead of instruction to continue with screw calibration.