REMEDY:
Turn the simulator off and allow solution to cool. Turn the simulator on. If the
condition continues to occur, simulator requires service*.
Remove the source of radio interference or change location of the simulator. Reset the simulator.
REMEDY
: Turn “off” – wait 3-5 seconds, then turn “on”.
REMEDY:
Reset the electronics. (Turn “off” for 3-5 seconds, then turn “on”.) If the simulator
does not heat, the heating element may be defective. Requires service*
The temperature of the solution is above
34.2
q
C or radio frequency interference
(RFI) has been detected.
Electronics need to be reset.
Excessive amount of time to obtain
operating temperature (longer than 15
minutes). The heating element may be
open or the solution was too cold.
MODEL 2100 REAR VIEW
!
CAUTION
REMOVE SOLUTION BEFORE
MOVING INSTRUMENT
REPLACE FUSE WITH BUSS TYPE GMW
1amp @ 125 vac Fast Acting