Emergency Locking
To lock the East/West sensor component, physically remove the whole locking
assembly from the frame by by removing the three screws arrowed in the photo
below:
It should be possible to turn the cam with a small pair of pliers applied to the
coupling sleeve on the gearbox output shaft. Do not apply pliers to the cam itself as
you risk damaging the surface. If the cam cannot be turned, release the single motor
clamp screw so that the motor and gearbox assembly can rotate as well. Turn the
cam until its crown faces the point of contact with the spring, then remount the
assembly. The mass will be locked as the screws are tightened.
Once the East/West sensor component is locked, remove the whole component by
following the procedure below:
14
Issue B - January, 2017