Fitting of End stops
End stops must always be used at the ends of rails,
whether free hanging, or butting up to a wall or other
elements .
• The End stop is placed in the rail before it is fitted .
• Before the End stop is placed in the rail, check that
the flange nuts are positioned correctly .
• When the rail system has been fully installed, place the
End stops in the correct position at the ends of the rails .
• If there is a risk that the End stop may fall out, a hole of
4 .5 mm (
1
/
4
") is drilled through the rail – not through the
angle bracket of the End stop .
• Hammer the accompanying spring pin into the End stop
and rail, flush to the metal plate .
• When dismantling the End stop the screws must be all
out, subsequently the angle profile including cylinder pin
can be tipped out, hence the pin is still attached to the
angle profile .
Adjustment of End stop at underhung traverse rail
(To avoid hoist for hitting the wall)
Hoist X=minimums measure
GH3
Trolley length
170 mm
(6
3
/
4
")
x = 100 mm
(4")
GH3+ Trolley length
210 mm
(8
1
/
2
")
x = 90 mm (3
1
/
2
")
GH1
Trolley length
170 mm
(6
3
/
4
")
x = 100 mm
(4")
GH1Q Trolley length
130 mm
(5
1
/
4
")
x = 125 mm
(5")
GH1F Trolley length
166/50 mm (6
3
/
4
" / 2")
x = 160 mm (6
1
/
2
")
End stop
x = min. distance to wall
Summary of Contents for GH1 F
Page 20: ... Guldmann GB US 09 2016 550376_3 22 GENERAL INFORMATION ...
Page 151: ... Guldmann GB US 09 2016 550376_3 153 TECHNICAL DRAWINGS Jumper settings GH3 GH3 Twin ...
Page 158: ... Guldmann GB US 09 2016 550376_3 160 TECHNICAL DRAWINGS ...
Page 164: ... Guldmann GB US 09 2016 550376_3 166 NOTES ...
Page 165: ... Guldmann GB US 09 2016 550376_3 167 NOTES ...
Page 166: ... Guldmann GB US 09 2016 550376_3 168 NOTES ...