Guildline 6636 Operator'S Manual Download Page 17

 

Section 4

 

 
OM6636-B-00 
12 February 2015 

4-1 

4.  MAINTENANCE 

4.1.

 

PREVENTIVE MAINTENANCE 

4.1.1.

 

Cleaning 

The instrument requires very little maintenance other than periodic dusting of the outside 
of  the  enclosure.    Use  alcohol  to  clean  around  the  standard  resistor  terminals  to  avoid 
formation of leakage paths. 

4.1.2.

 

Temperature 

Check  the  internal  temperature  with  the  imbedded  PRT  before  using  the  resistors  as 
references and also periodically. 

Compare the calculated temperature with the value given in the Report of Calibration. 

4.1.3.

 

Resistor Stability 

Resistor stability is determined by comparison against known reference resistors. 

4.2.

 

CERTIFICATION OF RESISTANCES 

If  the  resistance  of  the  standard  resistors  is  to  be  determined  by  direct  comparison  with  a 
standards laboratory, the comparison should be made with the resistors inside the enclosure at the 
operating temperature.  After transportation, or if the resistors’ temperature stabilized enclosure 
has been allowed to cool, the resistors should be allowed to stabilize at the operating temperature 
for at least 36 hours before measurement or before use. 

4.3.

 

CORRECTIVE MAINTENANCE 

Symptoms  appear  at  the  beginning  of  each  of  the  following  paragraphs  as  a  guide  to  detailed 
fault location.  Refer to Section 5 of this manual for the PRINCIPLES OF OPERATION. 

4.3.1.

 

Preliminary Checks 

Check POWER indicator - This should be lit when the instrument is connected to the AC 
line and the power ON/OFF switch in set to “ON”.  If not, check that both of the other two 
HEATER  and  OVERTEMP  indicators  are  also  not  lit.    If  they  are  both  not  lit,  then  a 
general loss of power is indicated.  If either indicator is lit, then it indicates that only the 
power indicator circuit is defective.  In addition, check the power outlet.  The fuse in the 
Power Entry module may be blown.  It may have been a defective fuse. 

Check the HEATER indicator - It should indicate the cycling on and off of the heater. 

Check  the  OVERTEMP  indicator  -  If  it  is  lit,  then  the  temperature inside the enclosure 

Summary of Contents for 6636

Page 1: ...to Guildline Instruments Limited They are to be used only as a guide to the operation and maintenance of the equipment with which this manual was issued and may not be duplicated or transmitted by an...

Page 2: ......

Page 3: ...line cord supplied or a proper line cord specified for the country of use Operating this instrument without a proper grounded line connection can result in electrical shock hazard 4 Observe all conne...

Page 4: ......

Page 5: ...ine Input Connector 1 3 1 4 2 Resistor Terminals 1 3 1 4 3 Guard Terminal 1 3 1 4 4 Chassis Ground Terminal 1 3 2 SPECIFICATIONS 2 1 2 1 GENERAL SPECIFICATIONS 2 1 2 2 RESISTANCE SPECIFICATION 2 2 3 O...

Page 6: ...iminary Checks 4 1 4 3 2 Temperature Control 4 2 4 3 3 Over Temperature 4 4 5 PRINCIPLES OF OPERATION 5 1 5 1 INTRODUCTION 5 1 5 2 TEMPERATURE SENSITIVE OSCILLATOR 5 2 5 3 AMPLIFIER AND DETECTOR 5 2 5...

Page 7: ...OF FIGURES FIGURE 1 1 FRONT VIEW 1 2 FIGURE 1 2 REAR VIEW 1 3 FIGURE 5 1 HEATER CONTROL UNIT 5 1 FIGURE 5 2 TEMPERATURE SENSITIVE OSCILLATOR 5 2 FIGURE 5 3 AMPLIFIER AND DETECTOR 5 2 FIGURE 5 4 HEATER...

Page 8: ...OM6636 B 00 12 February 2015 iv LIST OF TABLES TABLE 2 1 GENERAL SPECIFICATIONS 2 1 TABLE 2 2 RESISTANCE SPECIFICATION 2 2 TABLE 4 1 CONTROL CIRCUIT BOARD TESTING 4 2 TABLE 4 2 TRANSISTOR VOLTAGES 4 4...

Page 9: ...l shock and vibration in commercial carrier shipment To protect the standard resistors from overheating due to a malfunction of the temperature control there is an overriding over temperature control...

Page 10: ...excess of the operating temperature or in the event of control failure 1 3 4 PRT The platinum resistance thermometer imbedded within the temperature controlled chamber is configured as a four termina...

Page 11: ...esistor Terminals There are a set of two terminal input outputs for each installed resistance element Each set consists of two type N connectors labelled with the nominal value of each of the standard...

Page 12: ...Section 1 OM6636 B 00 12 February 2015 1 4...

Page 13: ...resistors 28 0 003 C 82 4 0 005 F Temperature Stability 1 year exclusive of the self heating effects of the resistors 28 0 01 C 82 4 0 02 F Cold Power on Stabilization Time to Specified Accuracy 36 ho...

Page 14: ...0 10 T 3500 3000 750 25 1000 100 T 6000 5000 1000 35 1000 Table 2 2 Resistance Specification Notes 1 Nominal initial tolerance is defined as the maximum variation of resistance mean values as initiall...

Page 15: ...on the rear panel as shown in Figure 3 1 Place the instrument on a vibration free bench or install it in an equipment rack as required Connect the line cord to the appropriate line power outlet and sw...

Page 16: ...the following equation C R R t t 392 0 0 Where R0 is the previously measured resistance value 3 2 3 Power Entry Selection Check to see that the fuses inserted in the receptacle correspond to the corr...

Page 17: ...r if the resistors temperature stabilized enclosure has been allowed to cool the resistors should be allowed to stabilize at the operating temperature for at least 36 hours before measurement or befor...

Page 18: ...rcuit If the indicator lights then check the resistance of Rl and the thermistor pair RT1 RT2 Rl should be 3 k to 5 k RT1 RT2 is equal to Rl at 30 C and is larger than R1 if the temperature is less th...

Page 19: ...4 T101 C102 R103 R125 R104 R110 TP1 If oscillation is observed then check between TP2 collector Q103 and Pl 7 The waveform should be as shown If not then check the following components Q103 C106 R117...

Page 20: ...mperature sensitive switch S2 is open If these conditions are satisfied then replace fuse F1 and check that the HEATER indicator is lit Monitor the increase in temperature with the PRT If the lamp doe...

Page 21: ...re consists of the following functional blocks a Temperature Control consisting of a sensor oscillator amplifier detector and heater b Over temperature Circuit c Power Supply Figure 5 1 HEATER CONTROL...

Page 22: ...utput to the base of Q101 increases As soon as this voltage exceeds half the transformer output the circuit gain becomes greater than one and oscillation occurs 5 3 AMPLIFIER AND DETECTOR Figure 5 3 A...

Page 23: ...ols the heater current The resistors R119 R121 and R122 ensure that Q1 is not conducting when there is no output from the oscillator 5 5 OVERTEMPERATURE CIRCUIT If the temperature in the enclosure exc...

Page 24: ...of the instrument from 100 120 VAC or 220 240 VAC 50 or 60 Hz and feeds a full wave rectifier The heater is fed from the unregulated output and the control circuit is fed from a regulator consisting...

Reviews: