Section 3
OM6622T-A-00
30 January
2018
3-1
3. QUICK MEASUREMENT GUIDE
3.1.
MEASUREMENT OPERATION GUIDE
This section contains information regarding the measurement of resistors in the range of 1m ohm
to 1G ohms. This guide is provided for manual front panel control of the 6622T bridge. An
operation guide for full automatic control of the 6622T bridge, through the terminal of the system
computer, is provided in the BridgeWorks Software User’s Manual. A typical example of a
resistance measurement response versus test sample is illustrated in Figure 3-1.
Figure 3-1 : Resistance Measurement Response Versus Test Sample
NOTE:
The last 35 samples of a total of 150 measurement samples should be used to
determine the mean measurement value and standard deviation of the sample
set, to allow sufficient stabilization time to achieve the specified uncertainty for
the model 6622T Bridge.