Chapter 10: Configuring a 4 - 20mA Zone Type
NOTE:
The span entered by the user is not the
m
but rather the ratio between the normal curve and
the correction.
Setting the Calibration Zero and Calibration Span
Setting the calibration zero and calibration span ensures accuracy in the values being monitored by the
system in the event that an input is not calibrated.
What You Should Know
The calibration zero and calibration span only need to be entered if the input calibration is incorrect.
1. In edit mode, under the
General
tab, click on the edit field beside the
Cal. Zero
button.
2. Enter the zero calibration.
3. Click on the edit field next to the
Cal. Span
button.
4. Enter the calibration span.
How it Works: Calibration Zero and Calibration Span
Calibrating the zero and span errors allow the installer to correct the output of a sensor. A sensor output is
viewed as a linear function y = mx + b
Sensors that are not calibrated can generate span or zero errors as illustrated in the figures below:
895-00677
Agri Alert 128 Touch
73
Summary of Contents for 890-00584
Page 14: ...NOTES 14 895 00677 Agri Alert 128 Touch...
Page 20: ...NOTES 20 895 00677 Agri Alert 128 Touch...
Page 30: ...NOTES 30 895 00677 Agri Alert 128 Touch...
Page 38: ...NOTES 38 895 00677 Agri Alert 128 Touch...
Page 44: ...NOTES 44 895 00677 Agri Alert 128 Touch...
Page 64: ...Chapter 9 Configuring a 0 5 Volt Zone Type 64 895 00677 Agri Alert 128 Touch...
Page 66: ...NOTES 66 895 00677 Agri Alert 128 Touch...
Page 82: ...NOTES 82 895 00677 Agri Alert 128 Touch...
Page 88: ...NOTES 88 895 00677 Agri Alert 128 Touch...
Page 92: ...NOTES 92 895 00677 Agri Alert 128 Touch...
Page 96: ...NOTES 96 895 00677 Agri Alert 128 Touch...
Page 100: ...NOTES 100 895 00677 Agri Alert 128 Touch...
Page 102: ...NOTES 102 895 00677 Agri Alert 128 Touch...
Page 106: ...NOTES 106 895 00677 Agri Alert 128 Touch...
Page 107: ...F Wiring Diagram with Terminals 895 00677 Agri Alert 128 Touch 107...
Page 110: ...NOTES 110 895 00677 Agri Alert 128 Touch...
Page 118: ......