23
Instructions manual |
www.grupotemper.com
KMD 06
Digital multimeter
Technical specifications
Accuracy: ±(% of r digits) at 18ºC ~ 28ºC with relative humidity of
<80%; guaranteed for a period of one year.
DC Voltage
Range
Resolution
Accuracy
200mV
0.1mV
±(0.5% of r 2 digits)
2V
0.001V
20V
0.01V
200V
0.1V
600V
1V
±(0.8% of r 5 digits)
Max. input voltage: 600V DC or AC rms.
AC voltage
Range
Resolution
Accuracy
200V
0.1V
±(1.0% of r 5 digits)
600V
1V
Max. input voltage: 600V DC or AC rms.
Frequency response: 40 ~ 400Hz, sine wave rms (avg. response)
Resistance
Range
Resolution
Accuracy
200Ω
0.1Ω
±(0.8% of r 4 digits)
2kΩ
0.001kΩ
20kΩ
0.01kΩ
200kΩ
0.1kΩ
2MΩ
0.001MΩ
Overload Protection: 250V DC or AC (RMS).