24
IND 220
6.3.
Error messages on the display
Display and
control panel
Cause
Remedy
•
E000
Heating tower overheated
Wait until the device has lowered the tempe-
rature and start again when the tower has
cooled down.
•
E001
Temperature probe does
not record an increase of
1 °C every 60 seconds
Check the temperature, material and dimen-
sions of the workpiece (→chap. 3.4
Technical
data). If these are not within the parameters,
the workpiece will not reach the programmed
temperature and will stabilise at a tempe-
rature below.
Check the position of the temperature probe.
It must be placed on the inner ring of the
workpiece (→chap. 5.4
Position temperature
probe).