Ground Zero ZeroStat MTR-1000 Manual Download Page 3

Ground Zero Electrostatics

MTR-1000 Megaohmmeter

th

4916 26  Street West, Suite 100, Bradenton, Florida 34207

  877-

 (463-1376), 

  941-751-7581 

 1-941-751-7586

 

[email protected]

Toll Free

GND-ZERO

Direct Phone

Fax

Email
Http://www.gndzero.com

Surface Resistance Measurements (RTT)

Surface-To-Ground Resistance 
Measurement (RTG)

Calibration & Maintenance

This procedure which complies with EOS/ESD-S4.1 measures resistance 
between two points independent of a groundable point.  Procedures vary 
regarding sample preparation, probe preparation, and spacing of the 5 
pound probes.  Select and read the correct test procedure for the desired 
measurements.

A.

  Connect the monaural plug ends of the test leads into the 3.5 mm jacks of 

the meter.  Connect the banana plugs of the test coil cords into the 2.5 inch 
disc ends of both 5 pound probes.

B.

  Place both probes on the material according to test procedures.

C.

  Move switch to desired test voltage position, either 10 or 100v.

D. 

 Press and hold the test button until power is applied to the meter and a 

value is displayed.  Keep the button depressed with sufficient force until the 
electrical resistance, relative humidity, and temperature readings are 
displayed on the meter screen.

This procedure measures the surface resistance between a ground point on 
the material surface and specific positions on the material being tested.  This 
procedure complies with the EOS/ESD-S4.1 standard.

A.

  Meter Set-Up. With both test leads connected to the meter, attach the 

alligator clip to one lead and the other to the 2.5" disc end of one 5 pound 
probe.

B.

  Attach the alligator clip to a known ground point.   Position the probe on 

the surface to be tested in accordance with the desired test procedure.

C. 

 Press the test button until the resistivity, humidity, and temperature test 

values are displayed on the LCD screen.  These readings will conform to:   
EIA, EOS/ESD, ANSI, IEC-93 CECC, and ASTM test procedures.  When 
performing tests, especially with high resistance materials, be sure the test 
lead wires do not touch or overlap and that your hands are not in contact with 
the probes or wires during the actual testing of the materials.  This will ensure 
accurate readings.

®

The ZeroStat MTR-1000 requires no service or maintenance except for an 
occasional cleaning of the rubber on the probes.  A mild soap and water 
solution will remove dirt or other harmful contaminants from  both the rubber 
probes and the meter case.  Harsher solvents will affect the rubber probes 
and therefore should not be used.  When the meter is not used for extended 
periods of time, remove the battery to prevent damage and/or leakage.  The 

®

ZeroStat MTR-1000 is calibrated to NIST traceable standards at the factory.  
The calibration is done by using a computer program and 1% resistors.  This 
newly improved method assures many years of long life and accurate 
readings.  Attaching 1% resistors to the probes or test leads will verify this 
accuracy.  

Figure 4 - ZeroStat MTR-1000 Point-to-Point Test

Figure 5 - ZeroStat MTR-1000 Point-to-Ground Test

Figure 6 - ZeroStat CB-9900 Calibration Box

Summary of Contents for ZeroStat MTR-1000

Page 1: ... S11 11 All materials must be tested on an insulated surface to avoid misleading measurements It is possible to measure down through the surface and not along the surface This is especially true with non homogeneous or multilayer materials It is possible to actually measure down through the dissipative surface layer and along the inner conductive surface and back up through the dissipative layer T...

Page 2: ...ment This compartment houses a supplied 9 volt battery which must be installed prior to use G Parallel Test Probes These probes are made from a highly conductive low durometer elastomer Care should be taken to avoid harsh solvents and extreme abrasion Occasional cleaning with a mild soap and water solution will extend the life of the probe feet If damage does occur these probes are replaceable for...

Page 3: ...h both test leads connected to the meter attach the alligator clip to one lead and the other to the 2 5 disc end of one 5 pound probe B Attach the alligator clip to a known ground point Position the probe on the surface to be tested in accordance with the desired test procedure C Press the test button until the resistivity humidity and temperature test values are displayed on the LCD screen These ...

Page 4: ...g probe conforming to EOS ESD ANSI and CECC standards optional Enclosed with these probes are two 6 foot 183 cm coil cord test leads Momentary on press button Warranty Exclusion Limit of Warranty The foregoing express warranty is made in lieu of other product warranties express and implied including merchantability and fitness for a particular purpose which are specifically on indirectly disclaime...

Reviews: