GREISINGER GMH 3691 Operating Manual Download Page 5

H36.0.12.6C-09 

Operating Manual GMH 3691 

page 5 of 12

 

 

5.2  Pushbuttons 

 

 

On/off key 

 

 

min/max when taking measurements: 

press shortly: 

min. or max. measuring value will be 
displayed 

press for 1 sec.:  the min. or max. value will be deleted 

up/down for configuration:

 

to enter values, or change settings 

 

CAL

:  

press shortly: 

show state of the electrode 

press for 2 sec.:  start oxygen calibration 

 

Set/Menu: 

press (Set) shortly:  change between  

oxygen concentration[%] and  
oxygen partial pressure [hPa]. 

press (Menu) for 2 sec.: configuration will activated 

 

Store/Quit: 

Measurement: 

Hold current measuring value (’HLD’ in 
display) 

Set/Menu: 

Acknowledge setting, return to measuring 

5.3  Connections 

 

Interface

: connection for electrically isolated interface adapter  

(accessory: GRS 3100, GRS3105 or USB3100)

 

Connection for oxygen sensor  

with integrated temperature probe 

 

The mains socket is located at the left side of the instrument.

 

 

Configuration 

For configuration of the device press 

 -key for 2 seconds: the main menu of the configuration will be called up. 

Use key 

 to select a sub-menu, use the key 

 to actually go into the selected sub-menu and to change parame-

ters. 

Use the keys 

 and 

 to set the individual value for the parameter. Press the key 

 again to memorize the 

changes ant to change to the main menu. Use key 

 to leave the configuration. 

 

’Pressure ABs.’: Input of Absolute Pressure

 

 

500 … 2000 hPa 

abs. 

The calculated oxygen values will refer to the en-
tered absolute.

 

 

’Offset’:  Zero Point Displacement of the Temperature

 

 

-3.0 °C ... 3.0 °C 

or 

-5.4 °F ... 5.4 °F 

The zero point of the measurement will be displaced 
by the value set to compensate for sensor and 
measuring device deviations. 

off: 

Zero point displacement is deactivated (=0.0°)

 

 

 

 

1

2

3

4

5

6

ON

OFF

max

CAL

Store

min

Set

Menu

Quit

   

2

Logg

A L

mg/l

ok

mmH g

hPa

SA L

hPa

%O

C A L

max

min

2

5

°C

   

2

Logg

A L

mg/l

ok

mmH g

hPa

SA L

hPa

%O

C A L

max

min

2

5

Summary of Contents for GMH 3691

Page 1: ...93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Hand held measuring device with alarm function for gaseous oxygen and temperature as of version V1 6 Operat...

Page 2: ...AGES 9 10 1 ERROR AND SYSTEM MESSAGES DURING OXYGEN CALIBRATION 9 11 THE SERIAL INTERFACE 10 11 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 10 12 SPECIFICATION 11 13 DISPOSAL INSTRUCTION 11...

Page 3: ...e operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using the device 1 Trouble free operation a...

Page 4: ...xygen and temperature values have been stable for a longer period of time CAL arrow indicates that an automatic oxygen calibration is carried out Alarm arrow indicates an alarm by blinking Warning tri...

Page 5: ...e probe The mains socket is located at the left side of the instrument 6 Configuration For configuration of the device press key for 2 seconds the main menu of the configuration will be called up Use...

Page 6: ...ower off delay in minutes Device will be automatically switched off as soon as this time has elapsed if no key is pressed no interface communication takes place automati cally deactivated for cyclic l...

Page 7: ...asuring please note When measuring gases please consider the following Calibration and measuring are depending of the absolute pressure at the sensor Therefore check absolute pressure before calibrati...

Page 8: ...lease note If You started editing the manual value once the change to air is no more possible 3 Calibration point 1 Pt 1 Expose sensor to the chosen gas and wait until temperatures of gas and sensor h...

Page 9: ...below the measuring range specified measuring value too low Sensor cable defective replace probe System fault switch on again if fault continues to exist de vice is damaged return to manufacturer for...

Page 10: ...ablished programming languages suitable for Windows XP Windows Vista Windows 7 Programming examples Visual Basic 4 0 Delphi 1 0 Testpoint 11 1 The following interface functions will be supported Chann...

Page 11: ...prox 2 mA Automatic off function Device will be automatically switched off if no key is pressed no interface communica tion takes place for the time of the power off delay The power off delay can be s...

Page 12: ...H36 0 12 6C 09 Operating Manual GMH 3691 page 12 of 12...

Page 13: ...ing chambers for cell cultures Monitoring of fermen ting process of fruits in fermentation plants etc Medicine Monitoring of oxygen concentration in respirators checking of breathing monitoring of oxy...

Reviews: