_____________________________________________________________________________
5
Start Operation
Connect sensor and turn on device via
key.
After segment test
the device displays some configuration:
• If the surface correction factor is adjusted, the display will show
„SCF.x“ and its value shortly.
After that, the device is ready for measuring.
6
Device Configuration
To change device settings, press
Set
(key 4) for 2 seconds. This will call the
configuration menu.
The parameters can be changed with
(key 2) or
(key 5).
Press
Set
again to go to the next setting.
Store
(key 6) finishes the configuration and returns to standard measuring operation.
Parameter
Values
Meaning
MENU
KEY
or
tYPE
ni.cr
J
S
T
N
B
E
Selection Thermocouple:
Type K: NiCr-Ni
Type J: Fe-CuNi
Type S: Pt10Rh-Pt
Type T: Cu-CuNi
Type N: NiCrSi-NiSi
Type B: Pt30Rh-Pt6Rh
Type E: NiCr-CuNi
rES
0.1°
Resolution
: 0,1°C or °F
1°
Resolution
: 1°C or °F
Unit
°C
Values in °C
°F
Values in °F
P.oFF
1-120
Auto Power Off
time in minutes
oFF
Auto Power Off
deactivated
Out
oFF
Function of the output: No output function, lowest power consumption
SEr
Output is serial interface
Adr.
01,11..91
Base
address
of interface (if Out = SEr)
OFS.1
-10.0°C..10.0°C or
-18.0°F..10.0°F
The
offset of sensor 1
will be displaced by this value to compensate for
deviations in the Sensor or in the measuring device.
oFF:
Zero point displacement is deactivated (=0.0°)
OFS.2
-10.0°C..10.0°C or
-18.0°F..10.0°F
The
offset of sensor 2
will be displaced by this value to compensate for
deviations in the Sensor or in the measuring device.
oFF:
Zero point displacement is deactivated (=0.0°)
SCL.1
-5.00..5.00 %
The measuring
scale of sensor 1
will be changed by this factor [%] to
compensate deviations of temperature probe or measuring device.
oFF:
Correction factor is deactivated (=0.000)
SCL.2
-5.00..5.00 %
The measuring
scale of sensor 2
will be changed by this factor [%] to
compensate deviations of temperature probe or measuring device.
oFF:
Correction factor is deactivated (=0.000)
SCF.1
0.950..1.200
Correction value for surface measurement of sensor 1.
(related to
ambient temperature
of the measuring device)
See chapter 9.3 9.3Correction for surface measuring
oFF:
Surface correction factor is deactivated (=1.000)
SCF.2
0.950..1.200
Correction value for surface measurement of sensor 1.
(related to
ambient temperature
of the measuring device) See chapter 9.3
oFF:
Surface correction factor is deactivated (=1.000)