GREISINGER GE 014 Operating Manual Download Page 3

Z03.0.0X.6C-11 

 Operating manual GE xxx serie 

page 3 of 12 

 

 

 

 

 

The manufacturer will assume no liability or warranty in case of usage for other purpose than the 
intended one, ignoring this manual, operating by unqualified staff as well as unauthorized 
modifications to the device. 

2.2  Safety guidelines 

This device has been designed and tested in accordance with the safety regulations for electronic 
devices. 
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety 
measures and special safety advises given in this manual will be adhered to when using the 
device. 

2.3  Safety signs and symbols 

The following signs in this document highlight warnings: 

 

Caution! 

This symbol warns of imminent danger, death, serious injuries and 

significant damage to property at non-observance. 

 

This symbol indicates danger for living tissue as well as a variety of materials, 
which can be damaged or destroyed when coming into contact with this 
chemical. Caustic effect, protective equipment required!

 

 

This Symbol indicates dangers to all living beings that may result in death or 
acute or chronic health hazards when inhaled, swallowed, or absorbed 
through the skin of this chemical. 

 

This Symbol indicates irritant substances that can cause inflammation on 
short-term, prolonged or repeated contact with the skin or mucous 
membranes. 

 

Attention!

 This symbol warns of possible dangers or dangerous situations 

that can provoke damage to the device or environment at non-observance. 

 

Note!

 This symbol point out processes which can indirectly influence 

operation, possibly cause incorrect measurement or provoke unforeseen 
reactions at non-observance. 

 

This symbol instructs the use of eye protection which protects the eyes from 
harmful influences when working with powerful light, UV radiation, laser, 
chemicals, dust, splinters or weather influences.

 

 

This symbol instructs the use of protective gloves which offer protection from 
mechanical, thermal, chemical, biological or electrical hazards.

 

 

 

Summary of Contents for GE 014

Page 1: ...Z03 0 0X 6C 11 Operating Manual pH electrodes GE xxx Keep for later use www greisinger de Companies Brands of the GHM English...

Page 2: ...uct for consulting in case of doubt 2 Safety 2 1 Intended Use The pH electrodes must be used only according to its intended purpose and under suitable conditions The electrodes must be operated with s...

Page 3: ...h can be damaged or destroyed when coming into contact with this chemical Caustic effect protective equipment required This Symbol indicates dangers to all living beings that may result in death or ac...

Page 4: ...bility cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using the device The electrodes contain 3 molar KCL GE103 1mol l...

Page 5: ...specification sheet of chemicals 3 Consider any statutory provisions guidelines and guidelines of chemicals when disposing This is also for accidentally spilled chemicals dried residues soiled rags or...

Page 6: ...he temperature avoid interference electrostatic charges etc keep plug contacts clean and dry prevent electrodes except special waterproof versions from extended immersion above the shaft calibrate the...

Page 7: ...ystems 3 4 Service life The service life of electrodes is normally at least 8 to 10 months When cared for properly this can usually increase to more than 2 years The actual life will vary depending on...

Page 8: ...to 30 C Below 5 C the electrode might be damaged because of freezing of the electrolyte We recommend to store the electrode vertically with the cable to the upper side The pH electrode should be arra...

Page 9: ...should be avoided as this might damage the electrode permanently Therefore a chemical cleansing is preferable Impurities Cleaners General residue Mild detergent Inorganic coatings common fluids for g...

Page 10: ...E 100 can be used Some different applications requires a special electrode GE 100 is a universal electrode with two ceramic diaphragms and liquid GE 101 is preferably used for small sample amounts It...

Page 11: ...disposal 6 1 Reshipment All devices returned to the manufacturer have to be free of any residual of measuring media and other hazardous substances Measuring residuals at housing or sensor may be a ri...

Page 12: ...Pt1000 sensor pH 0 14 0 60 C 100 S cm max 6 bar 3mol l KCl gel BNC and 4mm banana 2 m pressure resistant up to 6 bar with PG13 5 thread 2x ceramic cylinder PSU approx 12x120 mm GE 120 Cinch injection...