B-H90.0.01.DB2-2.1
25 / 30
G 1910-02
Error and system messages | 8
ERR.T
Temperature error
Permissible temper-
ature range during
charging exceeded
or undercut
Charging of battery only at 0
.. 40 °C
Bring product to room tem-
perature and restart charg-
ing process
SYS ERR
System error
Error in the product
Sensor defect
Switch product on/off
Replace rechargeable bat-
teries
Send in for repair
STAB
No measurement
change within 2
minutes
Product in extremely
constant environ-
ment
Sensor defect
Place the product in clean
outdoor air
Perform sensor adjustment
Send in for repair