Safety instructions:

This device has been designed and tested in accordance with the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety
advises given in this manual will be adhered to when using the device.
1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic

conditions than those stated under "Specification". If the device is transported from a cold to a warm environment condensation
may cause in a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature
before trying a new start-up.

2. General instructions and safety regulations for electric, light and heavy current plants, including domestic safety regulations (e.g.

VDE), have to be observed.

3. If device is to be connected to other devices (e.g. via PC) the circuitry has to be designed most carefully. Internal connection in third

party devices (e.g. connection GND and earth) may result in not-permissible voltages impairing or destroying the device or another
device connected.

4. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked accordingly to

avoid re-starting.
Operator safety may be a risk if:
- there is visible damage to the device
- the device is not working as specified
- the device has been stored under unsuitable conditions for a longer time.

 

In case of doubt, please return device to manufacturer for repair or maintenance.

5. 

Warning:

Do not use these product as safety or emergency stop devices, or in any other application where failure of the product could result
in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material damage.

Disposal instructions:

The device must not be disposed in the regular domestic waste.
Send the device directly to us (sufficiently stamped), if it should be disposed. We will dispose the device appropriate and
environmentally sound.

Assignment of elbow-type plug: (depending on type)

4-20mA (2-wire connection)

voltage, not electrically isolated

voltage, electrically isolated

(3-wire connection)

(3- resp. 4-wire connection)

(1 combined elbow type plug for humidity and

 

(2 elbow type plugs: right plug is for humidity

 temperature)

 left plug is for temperature)

  1

=  supply vUv 

(humidity)

1

(humidity)

1

(humidity or temp.)

2

=  GND / signal

  

(humidity)

2

(temperature)

2

=  signal -

(humidity or temp.)

3

= supply vUv  

(temperature)

3

=  supply vUv

3

=  supply vUv

 

(4)

=  GND / signal

  

(temperature)

(4)

=  supply voltage -Uv

(4)

=  supply voltage –Uv

(note:   and 2 are connected internally)

The type current or voltage output is set by works and cannot be changed

General installation instructions:

To mount the connection cable (2-, 3-, or 4-wire depending on type of device) the elbow-type plug screw has to be
loosened and the coupling insert has to be removed by means of a screw driver at the position indicated (arrow).
Pull out connection cable through PG glanding and connect to the loose coupling insert as described in the wiring
diagram. Replace loose coupling insert onto the pins at the transducer housing and turn cover cap with PG glanding in
the direction desired till it snaps on (4 different starting positions at 90° intervals). Re-tighten the screw at the angle plug.