Installation
16
Power Supply
– Main In
J11
Power Lead
J11-L
Neutral Live
J11-N
Protective earth
PE
Power Supply
– Unit 1 (A)
J9
Power Lead
J9-L
Neutral Live
J9-N
Protective earth
PE
Power Supply
– Unit 2 (B)
J10
Power Lead
J10-L
Neutral Live
J10-N
Protective earth
PE
Relay Output
– Common (A/B)
J3
R1
– Alarm High/Low O2 level
J3 Terminals 1-2
R2
– Alarm High High O2 level
J3 Terminals 3-4
R3
– System failure alarm
J3 Terminals 5-6
R4
– Not used by default
J3 Terminals 7-8
Relay Input from Unit 1 (A)
J5
R1
– Alarm High/Low O2 level
J5 Terminals 1-2
R2
– Alarm High High O2 level
J5 Terminals 3-4
R3
– System failure alarm
J5 Terminals 5-6
R4
– Not used by default
J5 Terminals 7-8
Relay Input from Unit 2 (B)
J4
R1
– Alarm High/Low O2 level
J4 Terminals 1-2
R2
– Alarm High High O2 level
J4 Terminals 3-4
R3
– System failure alarm
J4 Terminals 5-6
R4
– Not used by default
J4 Terminals 7-8
Analog Ouput
– Common (A/B)
J6
Analog output 1
– O2 %
J6 Terminals 1-2
Analog output 2
– optional
J6 Terminals 3-4
Analog Input from Unit 1 (A)
J8
Analog output 1
– O2 %
J8 Terminals 1-2
Analog output 2
– optional
J8 Terminals 3-4
Analog Input from Unit 2 (B)
J7
Analog output 1
– O2 %
J7 Terminals 1-2
Analog output 2
– optional
J7 Terminals 3-4
DIP Switches
Keyboard
A or B Power on
(Default setting)
DIP 1 & 2: ON, &
DIP 3 & 4: OFF
Connection to the two keys on the top cover.
A & B powered on
DIP 1 & 2: OFF, &
DIP 3 & 4: ON
Figure 3-6: Electrical connections of the selector box G36