40
5.
Enter Test Conditions:
2
3
4
5
1
1. Enter your name or other identification here.
2. The 42AP pistonphone supplies this information automatically. If the 42AP is not
connected to the COM port, enter the barometric pressure [hPa] manually, e.g.
“1005”.
3. Type in the temperature [°C], e.g. “22 “.
4. Type in the relative humidity [%], e.g. “42”.
5. To save your data, press “OK”.