24 | 12Bx Power Modules
In this example, the command turns on the red LED for 20 s,
followed by the green LED for 15 s, followed by the blue LED for 10 s
for the microphone on port 2.
RL
Description
Returns the current programmed reference level in volts when
entered with only the
--InputNumber
option. When a reference
voltage is included in the command, the reference level is set to the
defined voltage value.
Syntax
RL -
i
portnumber
or --
InputNumber
portnumber
[level]
Example
>MicPowerModuleApp.exe RL -i2 92.84
In this example, the stored reference level for the microphone on
port 2 is set to 92.84.
RT
Description
Returns the current programmed reference temperature in
degrees C when entered with only the
--InputNumber
option.
When a reference temperature is included in the command, the
reference level is set to the defined temperature value in degrees
C.
Syntax
RT -i
portnumber
| --InputNumber
portnumber
[
temperature
]
Options
-i
portnumber
or
--InputNumber
portnumber
The port number (1-99) of the connected microphone.
temperature
The temperature in degrees C.
Example
>MicPowerModuleApp.exe RT -i2 25
Summary of Contents for 12BA
Page 2: ......
Page 4: ......
Page 16: ...12 12Bx Power Modules Instruction Manual...
Page 38: ...34 12Bx Power Modules Instruction Manual...
Page 46: ...42 12Bx Power Modules Instruction Manual...