When measuring rotation speed with pulse input, why does it increase in 60times when the sampling
interval is 1 second?
The number of rotations is converted from the number of pulses measured at the sampling interval to
the number of rotations per minute.
When the sampling interval is 1 second, this is because the number of pulses counted within 1 second is
multiplied by 60 and displayed.
30
.
Is it possible to calculate the difference between the voltages measured on Ch. 1 and Ch. 2 and display it on
Ch. 3?
Also, can I use the value as an alarm setting?
Display is possible using the inter-channel calculation function. However, it is not possible to set alarms for the
channels using the calculation function.
31
.
Is it possible to set the upper limit of the scaling setting to a lower value than the lower limit?
It is possible.
The lower and upper limits of the scaling settings can be set to the opposite values such as lower limit > upper
limit.
32
.
How to change the range setting with channels that has already been set?
Hasn't the EU (scaling) been On?
If EU (scaling) is On, the range of the channel cannot be changed. Set EU (scaling) to Off and try change
the range.