4-17
CHAPTER 4 Settings and Measurement
Logic and Pulse settings
Makes settings related to digital input.
<For Pulse>
<For Logic>
(2)-7 Logic/Pulse
This is used to select the processing method for digital input.
Selection item
Description
Off
Digital input measurement is disabled.
Logic
Digital input is processed as logic signals.
Pulse
Digital input is processed as pulse signals.
(2)-8 Input
This is used to set the pulse measurement mode. This setting is available only if Pulse is selected in (2)-7.
Selection item
Description
Off
Pulse input measurement is disabled.
Revol.
The number of pulses per sample interval is counted and converted to the number of revolutions per
minute.
Counts
Captures the cumulative number of pulses for each sampling interval from the start of measurement.
Inst.
Captures the number of pulses for each sampling interval.
(2)-9 Filter
This is used to set the filter for input.
Selection item
Description
Off
Disables hardware filter.
On
Enables hardware filter. It is effective in a noisy environment.
The filter is approximately 30 Hz (-3 dB).
(2)-10 Pulse Slope
This is used to set the slope (direction) to count the number of pulses. This setting is available only if Pulse
is selected in (2)-7.
Selection item
Description
H
Counts the rising edges of pulses.
L
Counts the falling edges of pulses.