Chapter 5
26
Micro-Ion E Module Instruction Manual 354011, Rev. A
5.4
Error Responses
If a command cannot be processed, the module returns one of the error
responses listed in Table 5-1.
Table 5-1
RS–232 Error Responses
Message from
Module
Possible Causes
Solution
COMM ER
Parity error in command from host.
• Verify that parity from host matches parity
configuration for module.
• Non-programmable parity for module is no parity,
one stop bit (see page 44).
RANGE ER
SB command contains an invalid baud rate.
• Re-send SB command containing valid baud rate.
• Valid baud rates are 1200, 2400, 4800, 9600, or
NVRAM ER
Module cannot write data to non-volatile
memory (NOVRAM).
• Send FAC command to reset communication values
to factory defaults (see page 30).
• Re-send set, unlock, or toggle command.
• Return module to factory (see page 40).
SYNTX ER
• Command was improperly entered.
• Module does not recognize command.
• UNL command was sent when software
functions were already unlocked.
Re-enter command using proper syntax (see page 27).
9.99E+09
• Module cannot indicate a valid pressure.
• Micro-Ion E gauge is OFF.
• Send RS command to determine module status (see
• If necessary, replace gauge assembly (see page 40).
• If Micro-Ion E gauge is OFF, send IG1 command (see
LOCKED
Interface function is locked.
Send UNL or TLU command to unlock interface
functions (see page 27).
INVALID
• Micro-Ion E gauge is defective.
• IG1 command has been sent.
• If Micro-Ion E gauge is defective, replace gauge
• If IG1 command has been sent, send IG0 command
(see page 30).
Summary of Contents for Micro-Ion 354 E Series
Page 2: ......
Page 10: ...10 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 24: ...24 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 42: ...Chapter 6 42 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 46: ...46 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 52: ...52 Micro Ion E Module Instruction Manual 354011 Rev A...
Page 53: ......