5.0 Technical specifications
5.1 Input ranges sq16
Inputs
sq16
sq16plus
Resistance
2 wire
0 to 300,000R
0 to 300,000R
3 wire
-
0 to 10000R
4 wire
-
0 to 10000R
Voltage (Single
ended and Diff)
-0.3 to0.3V
-1.5 to 2.5V
-1.5 to 2.5V
-25 to 50V
-6 to 10V
-40 to 60V
Current
4 to 20mA
4 to 20mA
-30 to 30mA
-30 to 30mA
Thermistor
U
-
50⁰C to+150⁰C
-
50⁰C to+150⁰C
UU
-
50⁰C to+150⁰C
-
50⁰C to+150⁰C
Y
-
50⁰C to+150⁰C
-
50⁰C to+150⁰C
S
-3
0⁰C to+150⁰C
-3
0⁰C to+150⁰C
User
-
50⁰C to+150⁰C
-
50⁰C to+150⁰C
Thermocouple
B
250⁰C to +1820⁰C
250⁰C to +1820⁰C
C
0⁰C to +2320⁰C
0⁰C to +2320⁰C
D
0⁰C to +2320⁰C
0⁰C to +2320⁰C
J
-
200⁰C to +1200⁰C
-
200⁰C to +1200⁰C
K
-
200⁰C to +1372⁰C
-
200⁰C to +1372⁰C
N
-
200⁰C to +1300⁰C
-
200⁰C to +1300⁰C
R
-
50⁰C to +1768⁰C
-
50⁰C to +1768⁰C
S
-
50⁰C to +1768⁰C
-
50⁰C to +1768⁰C
T
-
200⁰C to +400⁰C
-
200⁰C to +400⁰C
Pt 100
2 wire IEC751
-
200⁰C to +850⁰C
-
200⁰C to +850⁰C
2 wire JIS1604
-
200⁰C to +850⁰C
-
200⁰C to +850⁰C
3 wire IEC751
-
-
200⁰C to +850⁰C
3 wire JIS1604
-
-
200⁰C to +850⁰C
4wire IEC751
-
-
200⁰C to +850⁰C
4 wire JIS1604
-
-
200⁰C to +850⁰C
Pt 1000
2 wire IEC751
-
200⁰C to +850⁰C
-
200⁰C to +850⁰C
3 wire IEC751
-
-
200⁰C to +850⁰C
4wire IEC751
-
-
200⁰C to +850⁰C