4. Operation of 

PMR-

10

0

4.1. 

Connect external power supply plug to the grounded mains.

4.2. 

The display will turn on with the upper line showing time, RPM and tilt angle set earlier (fig.1/

u

and the lower line showing current readings of the same parameters (fig.1/

v

).

How to set the necessary parameters.

Use the readings in the upper line of the display, while setting the necessary parameters.

4.3. 

Time set (Time  hr:min)

With the help of 

the 

p

” and “

q

“ TIME keys (Fig. 1/

w

) set the required working time interval in 

hours and minutes (increment - 1 min). If the button is pressed for longer time the speed of 
increment becomes bigger.

Note!

 If the operation time is not set and the timer indicator shows 0:00, pressing RUN key 

(Fig. 1/

z

will allow continuous use until the STOP

 key (Fig. 1/

{

) is pressed again.

4.4. 

Speed set (Speed RPM)

With the help of 

the 

p

” and “

q

“ SPEED keys (Fig. 1/

x

) set the required 

rocking

 intensity in 

revolutions per minute (increment 1 oscill/min). If the button is pressed for longer time the 
speed of increment becomes bigger. 

4.5. 

Tilt angle set (Angle Deg.)

With the help of 

the 

p

” and “

q

“ ANGLE keys (Fig. 1/

y

) set the necessary angle (increment 

1º). If the button is pressed for longer time the speed of increment becomes bigger. 

Run/Stop

OFF

30

90

10

90

10

SPEED

ANGLE

RUN

STOP

Time

(hr:min)

Speed

(RPM)

Angle
(Deg.)

Fig.1. Control panel

Platform Rocker PMR-100

Operating instructions

Version 

1.02

 - 

April 

20

1

3

Summary of Contents for Platform Rocker PMR-100

Page 1: ...Operating instructions Platform Rocker PMR 100 For version V 1GW ...

Page 2: ...Page 2 Platform Rocker PMR 100 Operating instructions Version 1 02 April 2013 ...

Page 3: ...3 Contents 1 Safety 4 2 General Information 5 3 Getting Started 6 4 Operation of PMR 100 7 5 Specifications 9 6 Guarantee and service 10 Platform Rocker PMR 100 Operating instructions Version 1 02 April 2013 ...

Page 4: ... do not use extension lead without grounding Only use the external power supply unit provided with this product Ensure that the external power supply connector is easily accessible during use Before moving disconnect from the mains To turn off the unit disconnect the external power supply from the mains outlet If liquid is spilled inside the unit disconnect it from the mains and have it checked by...

Page 5: ...protein electrophoresis Optimal for biomolecular hybridization on strips and for staining distaining procedures The PMR 100 can be installed inside an incubator ideal for cell growth in single use plastic bags working volumes of bags 5 litres Platform Rocker PMR 100 is equipped with a system of platform overload automatic detection In the case of overload an error message is shown on the unit s di...

Page 6: ...n of conformity 1 copy 3 3 Set up place the unit on an even firm stable horizontal surface connect the power cord to the external power supply plug the external power supply unit into the 12 V socket at the rear side of the unit 3 4 Platform installation install the platform on the unit and remove the rubber pad secure the platform on the supporting platform on top of the unit with the four suppli...

Page 7: ...of increment becomes bigger Note If the operation time is not set and the timer indicator shows 0 00 pressing RUN key Fig 1 z will allow continuous use until the STOP key Fig 1 is pressed again 4 4 Speed set Speed RPM With the help of the p and q SPEED keys Fig 1 x set the required rocking intensity in revolutions per minute increment 1 oscill min If the button is pressed for longer time the speed...

Page 8: ...ws the flashing reading OVERLOAD accompanied by the repetitive sound signal until the STOP key Fig 1 is pressed Eliminate the causes of overload before restarting operation see p 4 6 4 9 After the set time elapses the platform motion stops and the timer shows the flashing reading STOPaccompanied by the repetitive sound signal until the STOPkey is pressed 4 10 If necessary the unit can be stopped a...

Page 9: ...nge 10 Digital time setting 1 min 99 hrs 59 min non stop Time setting resolution 1 min Maximum continuous operation time 168 hrs Maximum loading 5 kg Platform working area 480 x 380 mm Dimensions 400x480x250 mm Input current power consumption 12V 1 1 A 13W External power supply input AC 100 240 V 50 60Hz output DC 12V Weight not more 13 kg Grant is committed to a continuous programme of improvemen...

Page 10: ...below return to our service department in the UK or in other countries our distributor 6 3 Cleaning Disinfection Cleaning liquids that do not contain concentrated organic solvents alkali or acid can be used for cleaning the device Standard ethanol 75 can be used for disinfection Before using any other cleaning or decontamination method check with the manufacturer or supplier to make sure that the ...

Page 11: ...Page 11 Platform Rocker PMR 100 Operating instructions Version 1 02 April 2013 ...

Page 12: ...atform Rocker PMR 100 1 02 Grant Instruments Cambridge Ltd Shepreth Cambridgeshire SG8 6GB UK Tel 44 0 1763 260811 Fax 44 0 1763 262410 Email scientificsales grantinstruments com www grantinstruments com ...

Reviews: