background image

Page 5

Centrifuge

 

LMC

-3000

 USA

Operating Instructions

Do not leave the operating unit unattended.

Do not operate the unit in environments with aggressive or explosive chemical mixtures.

Do not operate the unit if it is faulty or been incorrectly installed.

For indoor use only.

Do not use outside laboratory rooms.

Before using any cleaning or decontamination method except those recommended by the manufacturer, check 
with the manufacturer that the proposed method will not damage the equipment.

Do not make modifications to the design of the unit.

BIOLOGICAL SAFETY

Without bioseal the centrifuge is not a biosafety system in accordance to EN61010-2-20 and cannot be used for 
centrifuging hazardous materials contaminated with toxic, radioactive or pathogenic micro-organisms.

It is the user's responsibility to carry out appropriate decontamination if hazardous material is spilt on or inside 
the equipment.

Version 

6

 - 

December 

20

11

Summary of Contents for LMC-3000 USA

Page 1: ...Operating instructions Laboratory centrifuge LMC 3000 USA bio ...

Page 2: ...Page 2 Centrifuge LMC 3000 USA Operating Instructions Version 6 December 2011 ...

Page 3: ...rifuge LMC 3000 USA Operating Instructions Contents Version 6 December 2011 1 Safety 4 2 General Information 6 3 Getting started 7 4 Operation 9 5 Specifications 11 6 Maintenance 13 7 Declaration of Conformity 15 ...

Page 4: ...l ELECTRICALSAFETY Connect only to the mains with a voltage corresponding to that on the serial number label Ensure that the switch and plug are easily accessible during use Do not plug the unit into the main outlet without grounding and do not use extension lead without grounding Before moving the unit disconnect it from the mains To turn off the unit disconnect the power plug from the mains outl...

Page 5: ...d except those recommended by the manufacturer check with the manufacturer that the proposed method will not damage the equipment Do not make modifications to the design of the unit BIOLOGICALSAFETY Without bioseal the centrifuge is not a biosafety system in accordance to EN61010 2 20 and cannot be used for centrifuging hazardous materials contaminated with toxic radioactive or pathogenic micro or...

Page 6: ...l Information LMC 3000 USA is a modern benchtop laboratory centrifuge useful for sedimentation of cells bacteria yeast formed blood elements It provides operation with tubes and microtest plates LMC 3000 USA is designed for safe work metal protecting housing easy maintenance and wide application range in biochemical industrial and other type laboratories Version 6 December 2011 ...

Page 7: ...Rotor R 2 on request Rotor support stand RR U on request Adapter sets BN 13 75 BN 13 100 and BN 16 100 for R 12 15 on request 3 3 Set up place the centrifuge on the even stable and clean surface plug the mains power cord into the socket on the rear and position the centrifuge so that there is easy access to the power switch and connector it is necessary to observe the safety area of 300 mm around ...

Page 8: ... 3 4 Rotor replacement Hold the rotor with one hand and with the help of the supplied wrench 13 mm turn fixation nut fig 2 1 counter clockwise to release the rotor Replace the rotor and secure the new rotor carefully and turning the fixation nut tightly 1 Fig 2 Rotor replacement ...

Page 9: ...se occurs during acceleration which can be due to improper rotor fixation 4 1 Check the power cord for any signs of damage and replace if necessary Connect the power cord to the mains outlet Set the power switch in position I ON on the rear side 4 2 The centrifuge turns on and lid opens automatically The display shows the following readouts previously set time and speed in the upper line Set mode ...

Page 10: ... cause of imbalance 4 8 Centrifugation is stopped automatically after the set time elapses while display shows blinking indication STOP A sound signal is emitted after full stop of the rotor press RUN STOP button fig 3 3 to stop the signal 4 9 If necessary centrifugation can be stopped before the set time elapses by pressing RUN STOP button The set time interval will be shown on the display 4 10 P...

Page 11: ... not more 30 sec Display LCD Chamber working diameter 335 mm Dimensions 495x410x235 mm Weight not more 13 5 kg Speed range 100 3000 RPM increment 100 RPM Operating voltage 120 V 50 60 Hz 120 W 1 A Operating conditions The unit is designed for operation in cold rooms incubators and closed laboratory rooms at ambient temperature from 4 C to 40 C and maximum relative humidity 80 for temperatures up t...

Page 12: ...Page 12 Centrifuge LMC 3000 USA Operating Instructions Version 6 December 2011 Grant is committed to a continuous programme of improvement specifications may be changed without notice ...

Page 13: ... parts inside the centrifuge chamber Rotors and other accessories are not autoclavable 6 4 Replacement of fuses fig 4 Disconnect from the outlet Remove the power plug from the rear of the unit Pull out the fuse holder by applying leverage in recess A Remove the fuse from the holder Check and replace with the correct fuse if necessary for 230V T1Aor for 120V T2A 6 5 Routine safety tests If routine ...

Page 14: ...Page 14 Centrifuge LMC 3000 USA Operating Instructions Version 6 December 2011 ...

Page 15: ...Page 15 Centrifuge LMC 3000 USA Operating Instructions Version 6 December 2011 ...

Page 16: ...Grant Instruments Cambridge Ltd Shepreth Cambridgeshire SG8 6GB Tel 44 0 1763 260811 www grant co uk sales grant co uk Fax 44 0 1763 262410 Centrifuge LMC 3000 28713 6 ...

Reviews: