Data output RS-232C for PC or POS
This scale has a data output compatible with RS-232C and can communicate
with devices that use different protocols. To configure it, follow the following
procedure:
When the balance is in weighing mode, press the
. The screen will
display the message “PLUPro”. Press the
several times, until the screen
shows “Uart”.
Press four times
until the display shows “Adr” followed by a numeric
value. Select the value corresponding to the desired protocol in accordance
with the following information:
Protocol 0 -
Without data output
Protocol 1 -
Continuous output for remote display RD3.
Protocol 2 -
Output on request ($)
Protocol 3 -
Continuous output (for PC or compatible)
By pressing the
and
, you can choose the desired option, and after
press
to confirm.
To exit, press the
and the balance will return to normal mode.