Filtered A/D converter counts.
Unfiltered A/D converter counts.
20.
DIGITAL OUTPUTS
This section of the options menu allows the operation of the optional 3-relay
board available for the K3 indicator to be configured.
These 3 relay outputs can be used either for controlling an external signalling
of the high/low/ok limits in the running mode as a checkweigher (H-L mode)
or for the dosage of a product at two speeds. Moreover, it is possible to
select whether the dosage is performed for filling a receptacle on the scale
(loading) or for emptying a receptacle (unloading). The options in this section
of the menu are the following:
Running mode for controlling the optional relay board.
The relay outputs will be controlled by the H-L running
mode. The K1 relay closes when the weight on the
scale is greater than the limit
hIGh
. The K2 relay closes
when the weight is between the
LOW
and
hIGh
limits.
The K3 relay closes when the weight is below the
LOW
value.
The relay outputs will be controlled by the dosage
mode.
Allows each one of the 3 relays to be checked and
operated by the user.
Setting the upper and lower values for running the
“checkweigher” mode.
f-Cnt
R-Cnt
D Out
NODE
tEst
dOsI
ChECk
=
h-L