Gram FD 410 SERIES Operation Manual Download Page 13

13

Tare key

Power switch

CALIbRATION

Procedure

1.  Adjust the level gauge and preheat it for 30 minutes.

2.  Press 

 key.

3.  Press 

 key, the balance will read its full range (e.g.200), press 

 

key repeateadly, it reads the calibration point within the range (e.g.200, 
100, ...).

4.  Choose one calibration point and place the right weight on the pan. Press 

 key, the balance starts calibration, 

“ACAL” displayed.

5.  When the external calibration is 200.000g finished, the balance reads the 

value of the weight on the pan (e.g.200).

bASIC wEIghINg 

(Tare function)

Procedure

1.  Press the key 

, the balance will automatically proceed system 

initialization and deduct the tare weight.

2.  Place the container on the weighing pan.

3.  Press the 

 key.

Summary of Contents for FD 410 SERIES

Page 1: ...SERIES SERIE FD 410 OPERATION MANUAL MANUAL DE UTILIZACION...

Page 2: ...nnect the Power Adaptor 10 Install the External Facilities 10 Preheat Time 11 Adjust the Level Gauge 11 Display description 11 Keys description 12 Calibration 13 Basic weighing Tare function 13 Solid...

Page 3: ...ion with a computer 22 RS 232 Interface hardware 24 Routine maintenance and troubleshooting 25 Troubleshooting 25 Clearance 25 Clearing the stainless steel surface 25 Safety check 26 Troubleshooting 2...

Page 4: ...lentamiento 34 Nivelar la balanza 34 Descripci n del display 35 Descripci n de las teclas 36 Calibraci n 36 Pesaje b sico 37 Medici n de la densidad de l quidos y s lidos 37 Densidad de material s lid...

Page 5: ...n a un ordenador 47 Interface RS 232 48 Rutina de mantenimiento y soluci n de problemas 49 Soluci n de aver as 49 Limpieza 49 Limpieza de la superfiie de acero inoxidable 49 Revisi n de seguridad 50...

Page 6: ...tic interference will make the displayed value in this balance change Once the interference is dismissed the balance could operate in normal way Please avoid any liquid on the surface of the balance A...

Page 7: ...quality guarantee will automatically cease being effective If the balance does not work well please contact your local distributor or our customer service center Preparatory before using Storage This...

Page 8: ...pport Glass windshield Density accesories Power adapter User manual wind shield RS232 ports erial no power port the balance Plan view back view Top view adjustable feet pan support display level gauge...

Page 9: ...ibrations Excessive moisture Usage Conditions Please do not lay the balance in a quite damp palce for a long time If the balance is transferred from a lower temperature environment to a higher tempera...

Page 10: ...the beaker 5mm from the rim of it 4 Test the temperature of liquid with thermometer 5 Choose the right holder and hang it on the holder Connect the Power Adaptor Only the power adaptor offered by us...

Page 11: ...Adjust the level gauge of the balance the level gauge needs to be adjusted once its place is changed The two front bottom bolts are used to adjust the level gauge Spin the two bolts as shown in the fi...

Page 12: ...ion OK Reading shown is stable g Reading shown is given in grams g cm3 Reading shown is given in density units The balance is developing a stable reading UNABLE Error operation HHHHH The weight on the...

Page 13: ...00 100 4 Choose one calibration point and place the right weight on the pan Press key the balance starts calibration ACAL displayed 5 When the external calibration is 200 000g finished the balance rea...

Page 14: ...d material density Because the buoyancy of the liquid density depends on the temperature temperature varies 1 C the density of distilled water will change 0 02 In order to modify the effect of tempera...

Page 15: ...1 0 000g 6 Put the measured sample onto the upper plate 7 788g 7 Save sample s weight in the air 7 788g 8 Take off the sample 0 000g 9 Test sample s weight in the liquid 6 889g 10 Save sample s weigh...

Page 16: ...ate liquid CEnTIg ShaPE SOLID LIQUID 0 000g 4 Set weight s density assume the weight s density is 2 24g cm if the weight s density is unknown refer to A to test the weight s density two times repeat C...

Page 17: ...ld be set up to meet individual requirement 1 Press the key to enter setup program 2 Press the key to choose the parameter to be set up and then press the key to confirm 3 Anytime to exit the setup pr...

Page 18: ...menu press key to choose Line Feed InTEr press key for confirmation balance displays 0 SEC press key repeatedly to check the preset time interval then press key for confirmation If to choose 0 SEC it...

Page 19: ...g type then back to normal weighing mode Restoring factory setup 1 Press key for entering into menu of the balance PrInt displays 2 Press key repeatedly till InITIA displays 3 Press key for confirmati...

Page 20: ...tial liquid s measured temperature is higher than the displayed temperature press key to increase till the reading is same as the referential liquid s measured temperature 3 After setting up temperatu...

Page 21: ...Remarks a Set up liquid media s parameter as 1 distilled water while measuring solid s density b When measuring liquid s density user can choose weight as the measuring solid measure its density firs...

Page 22: ...99879 0 99877 0 99875 0 99873 0 99871 0 99870 0 99868 0 99866 0 99864 18 0 99862 0 99860 0 99859 0 99857 0 99855 0 99853 0 99851 0 99849 0 99847 0 99845 19 0 99843 0 99841 0 99839 0 99837 0 99835 0 9...

Page 23: ...J K L M A Signs field usually no display as a space when it is a positive number is displayed when it is a negative number B G Number and decimal field spaces are used when there are less than six dig...

Page 24: ...vice is as follows Pin description 2 TXD scale transmits data 3 RXD scale receives data 5 GRD signal ground Note Handshake signals such as clear to send CTS are not used The peripheral must have a m n...

Page 25: ...e Remove the shield ring and pan support with the weighing pan which helps avoid damage to the weighing system Clearing the stainless steel surface All the stainless Steel components need to clean reg...

Page 26: ...your Distributor The repairman must have had profesional training Troubleshooting DISPLAY REASON SOLUTION The balance could not get a stable reading or the transducer is damaged Contact your local se...

Page 27: ...vibration or air current or there is wrong connect with the pan Put the balance in another place Clear the wrong connect NO DISPLAY No work voltage no connection with transformer Check the power supp...

Page 28: ...to 40 C 41 F to 104 F The balance could assure the normal work 5 C 41 F to 40 C 104 F Technical parameters Model FD410 Capacity 410 g Readability 0 001 g Tare range 410 g Repeatibility 0 001 g Lowest...

Page 29: ...fect of the material During this period GRAM PRECISION SL covers the manpower and the spare parts necessary for the reparation of the scale This guarantee does not cover the failures caused by an inap...

Page 30: ...cias electromagn ticas excesivas estas pueden provocar una lectura del display err nea Una vez disminuidas estas interferencias electromagn ticas la balanza debe funcionar de manera normal Evitar cual...

Page 31: ...uministrado junto con la balanza Si la balanza no funciona correctamente por favor contacte con su distribuidor Antes de utilizar la balanza Entorno de la balanza Esta balanza requiere un entorno libr...

Page 32: ...trina de cristal Accesorios calculo densidad Adaptador de red Manual de instrucciones wind shield RS232 ports erial no power port the balance Plan view back view Top view adjustable feet pan support d...

Page 33: ...r la balanza en un lugar muy h medo durante mucho tiempo Si la balanza es transportada de un lugar con bajas temperaturas a un lugar con altas temperaturas por favor dejar la balanza en el nuevo lugar...

Page 34: ...impresora u ordenador Tiempo de Precalentamiento Para asegurar la exactitud en las lecturas la balanza debe ser precalentada durante 30 minutos antes de utilizarla Nivelar la balanza Nivelar la balanz...

Page 35: ...eto 4 Capacidad Resoluci n 5 Indicador OK lectura estable OK Lectura estable g Lectura en gramos g cm3 Lectura en densidad La balanza est desarrollando una lectura estable UNABLE Error de operaci n HH...

Page 36: ...de burbuja y precalentar la balanza durante 30 minutos 2 Pulsar la tecla para poner la lectura del display a cero 3 Pulsar la tecla el display mostrar la capacidad de la balanza por ejemplo 400 4 Pul...

Page 37: ...rimir el resultado pulsando la tecla Medici n de la densidad de l quidos y s lidos Formula Peso en el aire Peso en el aire Peso en el l quido Densidad del liquido Densidad de material s lido Debido a...

Page 38: ...lay pulsando la tecla Tara 0 000g 4 Predeterminar el tipo de la muestra en este caso s lido CEnTIg ShaPE SOLID 0 000g 5 Seleccionar la temperatura del l quido pulsar la tecla para incrementar el valor...

Page 39: ...sualizar el valor 1 221g cm 12 Salir del modo densidad 0 000g Densidad de l quido Paso Ejemplo de procedimiento Tecla Instrucci n Display 1 Instalar el kit de densidad ver p gina 31 2 Precalentar la b...

Page 40: ...epetir CEnTIg SHAPE COEFF 1 SET DP SET DP 1 00 2 24 2 24 parpadear 0 000g 6 Colocar el objeto en el plato superior 7 788g 7 Guardar el resultado del peso en el aire 7 788g 8 Retirar el peso del plato...

Page 41: ...0 0 99798 0 99795 0 99793 0 99791 0 99789 0 99786 0 99784 0 99782 22 0 99780 0 99777 0 99775 0 99773 0 99771 0 99768 0 99766 0 99764 0 99761 0 99759 23 0 99756 0 99754 0 99752 0 99749 0 99747 0 99744...

Page 42: ...ulsar la tecla para confirmar el display mostrar 0 SEC pulsar la tecla repetidamente para visualizar el intervalo de tiempo de impresi n deseado Pulsar la tecla para confirmar 6 Si se ha seleccionado...

Page 43: ...mar 5 Seguidamente el display visualizar PArITy configuraci n de la paridad 6 Pulsar la tecla para seleccionar el modo deseado el display muestra inicialmente el modo nOnE 7 Pulsar la tecla para selec...

Page 44: ...je Configuraci n de la temperatura 1 Pulsar la tecla 2 Pulsar la tecla hasta visualizar CEnTIg 3 Pulsar la tecla el display visualizar el valor de temperatura por defecto de 20 Si la temperatura media...

Page 45: ...ealizar es la densidad de un l quido pulsar la tecla el display mostrar LIQUID pulsar la tecla para confirmar Configuraci n de los par metros de la media 1 Pulsar la tecla 2 Pulsar la tecla repetidame...

Page 46: ...ciones opcionales Gancho para pesada hidrost tica En la parte inferior de la balanza hay un agujero para instalar el gancho de pesada opcional No est permitido utilizar el gancho de pesada inferior pa...

Page 47: ...de impresi n inmediatamente En respuesta a este comando la balanza transmitir cualquier n mero o mensaje que aparezca en el display El formato de salida es el siguiente A B C D E F G H I J K L M A Cam...

Page 48: ...a Interface RS 232 Esta balanza adopta la transmisi n y recepci n de l neas del RS 232 est ndar El formato de datos es 1 start bit 8 data bits include parity 1 stop bit Lasinstruccionesparaconectarlab...

Page 49: ...filtraci n de l quidos dentro de la balanza No utilizar en ning n caso limpiadores corrosivos disolventes Limpiar la balanza con un pa o de tacto suave Antes de limpiar la balanza retirar el plato el...

Page 50: ...co o con su Distribuidor para su reparaci n Soluci n de problemas DISPLAY MOTIVO SOLUCI N La balanza no puede obtener una lectura estable o la c lula de carga est averiada Contactar con su Distribuido...

Page 51: ...lato no est instalado correctamente Situar la balanza en un lugar m s apropiado Comprobar e instalar el plato adecuadamente NO DISPLAY No hay tensi n Comprobar el transformador de corriente de la bala...

Page 52: ...t cnicos Modelo FD410 Capacidad 410 g Resoluci n 0 001 g Rango de Tara 410 g Repetibilidad 0 001 g Lectura m nima de densidad 0 01 g Tiempo de respuesta 2 5 s Valor de la pesa de calibraci n 200 o 40...

Page 53: ...riodo de 1 a o a partir de la fecha de entrega Durante este periodo GRAM PRECISION SL se har cargo de la reparaci n de la balanza Esta garant a no incluye los da os ocasionados por uso indebido o sobr...

Page 54: ...54 Notes...

Page 55: ...55 Notes...

Page 56: ...Gram Precision S L Traves a Industrial 11 08907 Hospitalet de Llobregat Barcelona Spain Tel 34 902 206 000 34 93 300 33 32 Fax 34 93 300 66 98 comercial gram es www gram es 001 08072016...

Reviews: