20 |
BIOBASIC MANUAL
Sensor offset is used in cases where there are deviations in the
cabinet’s actual operation compared to control measurements by
independent temperature measuring systems.
The A-sensor is used to manage the cabinets refrigeration system, and
is also the reference sensor for the display and alarms.
The A-sensor is offset if the actual temperature in the cabinet does not
match the setpoint, despite taking the hysteresis into consideration.
Offset of A sensor is named “CA”.
Offsetting the A-sensor
Press and hold
for more than 5 seconds
Press
to select “CA”
Press
+
to offset the A-sensor
Press
to confirm the set value
The A-sensor is now offset, proceed to other parameters by pressing
, and then navigate by using
or
Leave the user menu by pressing
several times until
the cabinet temperature is shown in the display.
Sensor offset