GD-450-525-INS-LAB-RevA21 • Regency 450 / 525 User Manual
21
CHECK UPHOLSTERY
Check upholstery at least once a month. Inspect for rips, tears and worn spots. Ensure that
all upholstery-attaching screws are present, properly aligned, and well-secured. Use a Phillips
screwdriver to tighten upholstery mounting screws. Check attaching screws for sharp edges or
stripped screws, and replace if found.
WARNING: Worn or torn upholstery, or upholstery with loose hardware, must be replaced
immediately. It may not support body weight.
CHECK REAR WHEEL ADJUSTMENT
Inspect the wheels at least once a month. Check alignment by elevating the rear of the wheelchair
on a stable object until the rear wheels clear the ground, or placing the wheelchair upside down
in a stable position such that the wheels can spin freely. Spin the wheels; there should be no
wobble or sideplay, and the wheels should spin freely without binding. The bearings should be
clean and rotate smoothly. Check rim sideplay to verify that bearings are not too loose. If there
is a problem, contact your Graham-Field authorized distributor. If adjustment is required, this
should be done by your Graham-Field authorized distributor only.
CHECK ARMS
Check arms at least once a month. Inspect for sharp edges or cracks which could weaken the arm,
and replace if found. Ensure all attaching screws are present and tight. Use a Phillips screwdriver
to tighten armpad mounting screws. Check for burrs on the screw heads and replace if found.
Ensure that screws do not extend into the padding. Confirm that the posts at base of arm fit
correctly in the sockets, snug but not binding.
CHECK BACKPOSTS / PUSH HANDLES
Check backposts at least once a month. Ensure that all mounting hardware is securely fastened
and that the backposts are not bent or damaged. Contact your authorized distributor if you
observe cracking or peeling paint or plating.
CHECK FOOTRESTS & ELEVATING LEGRESTS
Check footrests and legrests at least once a month. Inspect the locking mechanisms to confirm
sure fit. Check for cracks, burrs, or sharp edges, and replace if found. Ensure that the footrest or
legrest will automatically lock securely in place and will not accidentally unlock.
CLEANING YOUR REGENCY 450 / 525
Wipe off base and seat frame frequently, at least once a week, using a soft cloth. Dry the
wheelchair immediately if exposed to moisture. Clean the frame every three months with a mild
soap and water solution. The frame does not need to be waxed. Do not use solvents, abrasive
waxes, caustic chemicals or spray silicone. Never use abrasive cleansers; they could scratch the
finish. Never use steam or high pressure cleaners. Clean upholstery and plastic components at
least once a month with a mild soap and water solution.