LCM-INS-LAB-RevD18 13
CLEANING AIR PUMP AND TUBES
1. Ensure pump is unplugged.
2. Wipe surface of pump and tubes with coarse cloth using an approved disinfectant germicide
mix to manufacturer’s instructions.
s
NOTICE: Do not flood any part of the pump with cleaning solution. Do not immerse the
pump in liquid.
s
NOTICE: Avoid spilling solutions on pump controls. If spills do occur, clean fluid from
pump wearing rubber gloves or while unit is unplugged to avoid any possibility of shock.
Once fluid is removed, check operation of components in area of spill. Fluids remaining on
controls can cause corrosion, which may cause components to fail or operate erratically,
possibly producing potential hazards for patient and staff.
3. Wipe off excess solution.
4. Disinfect pump and tubes with chlorine solution (mixed accordingly). Using a clean cloth,
wring out excess solution until cloth is damp. Wipe pump surface and tubes with damp cloth.
5. Allow to air dry.
Preventive maintenance for the Convertible Mattress consists of regular cleaning and an overall
system check-out to be performed at the intervals described below.
All components must be cleaned, disinfected and inspected after each patient’s use and before use
by a new patient. Always use standard precautions, treating all used equipment as contaminated.
Institutions should follow local protocols for cleaning and disinfection.
s
NOTICE: DO NOT wash Convertible Mattress under water pressure or steam clean.
INSPECTION / SYSTEM CHECK-OUT
Check each of the following before placing the Convertible Mattress with a new patient:
1. Check mattress surface for tears or cracking; do not use if tears or cracks are present.
2. Ensure mattress is free of stains and is not overly faded.
3. Ensure air inlet hoses and connectors on mattress and pump are clean and undamaged.
4. Ensure pump and power cord are clean and undamaged.
5. Ensure pump hanger brackets are secure and operate correctly.
6. Ensure OFF / ON Power switch and comfort control knob both operate correctly.
7. Attach pump to the blue Alternating Pressure hoses and power up to ensure there are no air
leaks.