Set Up
3A5411J
5
Set Up
1.
Install four, AA alkaline batteries (9) in
the battery holder (8). See
, page 14, Steps 1 - 6.
Register the TLM
It is recommended that the TLM be
registered prior to installation.
The TLM operating parameters are
controlled by the Pulse software and set up by
the System Administrator.
NOTE:
Manual TLM readings taken prior to
TLM configuration will be incorrect.
To register the TLM:
1.
On the Pulse web portal, set the Hub, or
the remote extender that the TLM
communicates with, to DISCOVERY
mode.
2.
Press and hold the Display button (A)
until the display (B) reads REGISTER,
then release (F
. 2). This may take up to
20 seconds.
NOTE:
A TLM will only display REGISTER if
a Hub, or remote extender, is within range
and in DISCOVERY mode.
3.
The display reads REGISTER OK (F
3, page 6) when the TLM is registered
with the Pulse software. This may take
up to 20 seconds.
If the TLM did not register with the Pulse
software, the error message REGISTER
E7 displays (see Step 1, F
Do not install or service this equipment
unless you are trained and qualified.
Installing and servicing this equipment
requires access to parts which may cause
fire, explosion, and serious injury if work is
not performed properly. Read
page 3.
NOTICE
Do not over tighten TLM into tank bung!
Over tightening can cause permanent
damage and result in inaccurate readings.
Do not use the TLM with pressurized
tanks.
Using the TLM in a pressurized tank
will damage the TLM.
Do not use thread sealant or adhesive.
Many of these products are chemically
incompatible with the Pulse/ABS plastic.
PTFE tape is acceptable.