Model 75-810 & 75-810-AR Noble Gas Purifier
Rev. 0 11/20
13
3
75-810 Specifications
Gases Purified
He, Ne, Ar, Kr, Xe
*Impurities Removed:
H
2
O, H
2
, O
2
, N
2
, CO, CO
2
, CH
4
Impurities Not Removed:
He, Ne, Ar, Kr, Xe
Operating Temperature:
400 °C + 25 °C
Temperature Sensor:
K-Type thermocouple is integrated in
the heater, monitor the cartridge
temperature by connection with a
flat pin connector to the receptacle
on the top panel
Maximum Recommended Operating Pressure: 17.0 atm (250 psig)
Suggested Maximum Flow Rates (ccpm):
Model 75-810: 1500
**Gas Connections:
1/4" VCR® Male
Leak to Atmosphere
<2 X
10
10
atm-cc/sec helium
***Lifetime Purifying He or Ar:
75-810: 1 yr. @ 200 ccpm
Power Requirements:
75-810: 80 Watts, 120 VAC
75-812:
80
Watts,
230
VAC
Circuit Breaker
In power switch, fast acting, 1A rating
Weight
(net):
3.5
lbs.
Dimensions:
5”D x 4”W x 10"H
*
Each impurity removed to < 10 ppb based on 10 ppm total inlet impurities at suggested
maximum flow rate
**
VCR® is a registered trademark of Swagelok Corporation
***
Lifetime based on suggested maximum flow rates; a higher flow rate will proportionally
decrease the lifetime while a lower flow rate will increase the lifetime. The lifetime is rates at
10 ppm total inlet impurities; higher or lower inlet impurity levels will decrease or increase the
lifetime proportionally. The purifier bed is considered an expendable item. System leaks, poor quality
gas, and improper handling will adversely affect its useful life.
Summary of Contents for 75-810-AR
Page 2: ...GOW MAC Instrument Co 2 ...
Page 6: ...GOW MAC Instrument Co 6 ...
Page 8: ...GOW MAC Instrument Co 8 ...
Page 10: ...GOW MAC Instrument Co 10 ...
Page 14: ...GOW MAC Instrument Co 14 ...
Page 16: ...GOW MAC Instrument Co 16 ...
Page 18: ...GOW MAC Instrument Co 18 ...
Page 22: ...GOW MAC Instrument Co 22 ...