ILIM, ILIM? Current Setting Limit Value
Settings
Function and manual operation:
See explanation on page 25.
Programming
Setting command:
IL
IM
v
Parameter:
v (value)
Parameter type:
Real number
Setting range:
0
v
ILIM
max
If the selected value is not a whole number multiple of the step
size, it is rounded off accordingly.
If necessary, the numeric value is rounded off once again for the
4-place digital display.
Default setting:
Nominal current (after RESET, *RST)
Example
(HP BASIC
):
OUTPUT 713; "ILIM 20"
{current setting range limit: 20 A}
Comment
ILIM may not be set to a value of less than ISET!
Any setting command to this effect is not executed.
Bit 1
would otherwise be set in event register B (
limit error
) and
bit 4
in the standard event register (
execution error
).
The ILIM function does not effect the adjustment of ISET by
means of a control signal via the analog interface.
Query
Displays the selected
final value
for
current setpoint
ISET
.
Programming
Setting command:
IL
IM
?
Response string:
ILIM
v
Length:
13 characters, fixed
Parameter
v (value)
Format
v
: +nn.nnnn
Example
(HP Basic)
:
OUTPUT 713;"
IL
IM
?
"
ENTER 713; A$
DISP A$
Display: ILIM
+20.0000
IMAX? – Maximum Measured Current Value
Function and manual operation:
See explanation in chapter 4.8.1.
Programming
Query command:
IM
AX
?
Response string:
IMAX
v
Length: 13
characters,
fixed
Parameter:
v (value)
Format: +nn.nnnn
Example
(HP Basic):
OUTPUT 713; "IMAX?"
ENTER 713; A$
DISP A$
Display:
IMAX +02.8550
IMIN? – Minimum Measured Current Value
Function and manual operation:
See explanation in chapter 4.8.1.
Programming
Query command:
IM
AX
?
Response string:
IMAX
v
Length: 13
characters,
fixed
Parameter:
v (value)
Format: +nn.nnnn
Example
(HP Basic):
OUTPUT 713; "IMIN?"
ENTER 713; A$
DISP A$
Display:
IMIN +02.7300
IOUT? – Querying the Momentary Current Value
Function and manual operation:
See explanation on page 23.
Programming
Query command:
IO
UT
?
Response string:
IOUT
v
Length: 13
characters,
fixed
Parameter:
v (value)
Format: +nn.nnnn
Example
(HP Basic):
OUTPUT 713; "IOUT?"
ENTER 713; A$
DISP A$
Display:
IOUT +02.7350
ISET, ISET? – Current Setpoint
Settings
Function and manual operation:
See explanation on page 21.
Programming
Setting command:
IS
ET
v
Parameter:
v (
value)
Parameter type:
Real number
Setting range
0
v
ILIM
If the programmed value is not a whole
number multiple of the step size, it is
rounded off in accordance with selected
setting resolution.
Device Type
Nom. current [A]
2
3
6
10
12
20
ILIM
max
[A]
2.0
3.0
6.0
10.0 12.0 20.0
Step size [mA]
1.0
1.0
1.0
1.0
1.0
1.0
Device Type
Nom. Current [A]
2
3
6
10
12
20
ISET
max
[A]
2.0
3.0
6.0
10.0 12.0 20.0
Step size [mA]
0.5
1.0
2.0
2.5
3.33
5.0