40
GMC-I Messtechnik GmbH
SbIt – Selecting the Number of Stop Bits
•
Either 1 or 2 stop bits can be used.
!
Settings at the SSP KONSTANTER and the controller must be
identical.
Settings
☞
Procedure
for selecting
Sbit
and setting the desired parameter
(see chapter 4.7):
!
Sbit
appears in the
buS
display along with the related
parameter (number of stop bits).
Figure 4.7.4 e
Path to Selection of the SbIt Text Parameter
4.8 Settings with the <SELECT> Key
The <
SELECT
> key is located on the front panel next to the
<
FUNCTION
> key. Depending upon settings in the
FUNCTION
menu,
it makes 3 different functions available:
4.8.1
In the Basic Function
Figure 4.8.1
Setting Basic Functions with the SELECT Key
Switching the 7-segment display from Uout and Iout to the
following measured values:
Pout
xxx.x
Max. output power
U_
xx.xx
Min. output voltage
U
xx.xx
Max. output voltage
I_
x.xxx
Min. output current
I
x.xxx
Max. output current
•
When the <
SELECT>
key is pressed the first time, the last
selected parameter name appears at the left-hand display,
and the respective parameter values appears at the right-
hand display.
•
When the <
SELECT>
key is pressed again, the next parameter
name with value is displayed, and so forth.
•
Display with the <
SELECT>
key is exited:
☞
By slightly turning rotary knob [5] or [8].
☞
By pressing the <
CE/LOCAL>
, <
SAVE>
, <
RCL>
or
<
FUNCTION> key.
4.8.2
During a Sequence Run and with Step-by-Step Control
The LOCKED/SEQ LED blinks when the
SEQUENCE
function is
active. The
SELECT
menu is expanded. It doesn’t matter whether
the
SEQUENCE
function is in the
RUN
or the
HOLD
status.
Pout
xxx.x
Max. output power
U_
xx.xx
Min. output voltage
U
xx.xx
Max. output voltage
I_
x.xxx
Min. output current
I
x.xxx
Max. output current
rcL
xxx
Current address, memory location number
rrEP
xxx
Number of remaining sequence repetitions
USEt
xx.xx
Programmed voltage value
ISEt
x.xxx
Programmed current value
tSEt
xx.xx
Programmed dwell time (tSEt or tdEF)
for current address, no remaining time
SSEt
txt
Output level for SSEt switching function
Otherwise, the same functions can be executed as listed in
chapter 4.8.1.
4.8.3
Display of Stored Data Upon Execution of <RCL>
Display of Data Stored to Setup Memory (1 ... 10) Upon Execution
of <RCL>
After executing the <RCL> command, selecting a memory
location (<
> or <
> key) and pressing the <ENTER> key, the
device is switched to the read-out mode. Values saved to setup
memory blink at the display. The basic settings stored to the
selected memory location can now be browsed by pressing the
<
SELECT
> key:
Uset
Iset
Voltage and current setpoints
tSEt
Programmed dwell time
SSEt
Signal output status
outP
Output switching status after power on
Ulim
Voltage setting limit
Ilim
Current setting limit
OVP
Overvoltage setpoint
OCP
Overcurrent protection
dLY
Overcurrent delay
UI_
Min-Max value recording
tdEF
Dwell time independent of memory location
Strt
Start address
StoP
Stop address
rEP
Number of repetitions
From the buS function group
*
ENTER
CE/LOCAL
FUNCTION
CE/LOCAL
Uout
Iout
FUNCTION
SELECT
SELECT
SELECT
SELECT
SELECT
SAVE
RCL
Uout/Iout
Uset/Iset
FUNCTION
MENU
SAVE
MENU
RCL
MENU
Abort
Abort
Abort
Abort
Abort