17.2.2 Selecting and Configuring the Integrated Test Sequence at the Test Instrument
All of the settings for the integrated test sequences can be
adjusted at the test instrument.
Select a switch position (A1 … A9), after which the start page for
the respective integrated test sequence is displayed (either the
default integrated test sequence or the previous setting).
The classification parameters (page 81) and the sequence param-
eters (page 87) can then be set.
Classification Parameters
DUT classification is set at the test instrument via classifi-
cation parameters (protection class, connection type,
measurement type etc.).
The test instrument attempts to automatically detect the set-
tings of certain classification parameters. In each case, an orange
border is displayed if this has been successful. Example for “test
socket connection type and protection class I”:
Automatic Detection Active for Protection Class
When connecting or disconnecting a DUT, the pro-
tection class can be changed without confirmation.
Automatic Detection Inactive for Protection Class
The test instrument retains the selected protection
class setting when a DUT is connected or discon-
nected.
Automatic setting of classification parameters must be checked
by the user. If they’re not recognized automatically, or if they’re
not recognized correctly, they have to be entered manually.
➭
Open the classification parameters menu by pressing
the key. All available parameters are displayed.
➭
Select the parameter to be changed with the softkey. A selec-
tion of possible settings appears.
Note
Descriptions of these parameters and their setting
options are shown in the following tables.
➭
Select the desired setting with the softkey. The classification
parameters menu is displayed again.
➭
Change additional settings if necessary.
➭
Confirm and save all of your changes with the checkmark.
The test menu’s initial page is displayed again.
Classification parameters can be changed very conveniently in the
case of test instruments with touchscreen (feature E01).
➭
The corresponding selection menu appears after touch click-
ing (briefly tapping) the respective classification parameters
window.
➭
The display is automatically returned to the start menu after
selecting the desired parameter.
Classification Parameters – VDE 0701-0702 /
ÖVE E 8701 / SNR 462638
1
These parameters must be entered manually if they’re not automatically de-
tected, or if they’re detected incorrectly.
2
The limit value for protective conductor resistance is determined on the ba-
sis of length and cross-section. Data remain in memory until a new entry is
made.
Classification Parameters – VDE 0701-0702-EDV /
ÖVE E 8701-EDV / SNR 462638-EDV
1
These parameters must be entered manually if they’re not automatically de-
tected, or if they’re detected incorrectly.
2
The limit value for protective conductor resistance is determined on the ba-
sis of length and cross-section. Data remain in memory until a new entry is
made.
Parameter
Setting Options / Meaning
1/2
Standard / Test Sequence
Currently selected standard with national designation
(refer to section 8.2 “Test Standards / Configuration of In-
tegrated Test Sequences”, for a list of standards available
for selection in the test instrument)
Protection class
1, 2
PCI, PCII, PCI+II, PCI+III, PCII+III, PCI+II+III
Connection type
1, 2
Test socket
Permanent connection
Adapter: AT16/32-DI adapter
Adapter: VL2E
Adapter: AT3-Adapter (feature I01)
Permanent connection: P1+P2 (only with feature H01)
2/2
Measurement type (MT)
1
Passive
Active
Auto-detection
Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & meas. type
Protection class only (PC)
Measurement type only (MT)
Off (= no auto-detection: all classification parameters
such as connection and protection class must be entered
manually)
Parameter
Setting Options / Meaning
1/2
Standard / Test Sequence
Currently selected standard with national designation
(refer to section 8.2 “Test Standards / Configuration of In-
tegrated Test Sequences”, for a list of standards available
for selection in the test instrument)
Protection class
1, 2
PCI, PCII, PCI+II, PCI+III, PCII+III, PCI+II+III
Connection type
1, 2
Test socket
Permanent connection
Adapter: AT16/32
Adapter: AT3-Adapter (feature I01)
2/2
Measurement type (MT)
1
Active
Auto-detection
Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & meas. type
Protection class only (PC)
Measurement type only (MT)
Off (= no auto-detection: all classification parameters
such as connection and protection class must be entered
manually)