GOSSEN-METRAWATT
25
Temperature ranges / Climatic class
Operating
temperature
–10 °C + 55 °C
Storage temperature –25 °C + 70 °C (without batteries)
Climatic class
2z/-10/55/70/75% with reference
to VDI/VDE 3540
Mechanical configuration
Protection type
Case IP 52
Probe IP 20
to EN 60529 / VDE 0 470
Dimensions 165 mm x 125 mm x 11O mm
without test leads
Weight
1.85 kg approx., including batteries
10
Maintenance
Attention!
!
Completely disconnect the tester from all external
circuits prior to replacing batteries, bulbs or fuses!
10.1 Messages on the LCD display
Messages that appear on the LCD display in place of the
digital display have the following meaning:
U LO
is show, when
• there is a blown or missing fuse (for ISO-
Ω
only)
• the battery voltage is too low
• the tester is defective
FUSE
is shown when pressing the measurement pushbutton
(8) in the
Ω
funktion, when
• there is a blown or missing fuse
OL
is shown, when
• there is an overrange condition
– – – –
is shown, when
• the tester is in „standby“ mode
10.2 Testing the LCD display
í
Turn the switch (6) from any position to another and
simultaneously press the pushbutton „RANGE
↑
“ (3) and
the pushbutton „RANGE
↓
“ (4). You thus activate the
LCD test.
í
Now briefly press the pushbutton „RANGE
↓
“ (4). All seg-
ments of the display, except for the warning triangle „V“
are presented on the LCD display (5).