11.3 Periodic Self-Test of the Connector Cable for Protective Conductor Continuity
Connect the probe cable (21) to a grounding contact which has been previously tested for
absence of voltage (e.g. at an electrical outlet), and which is connected to the protective
conductor in the connector cable, and set the NETZ-VDE switch (1) to “NETZ”. Then
measure protective conductor resistance as described in section 5.4. If an excessively high
protective conductor resistance value is displayed at the LCD panel or if overloading is
indicated (only “O.L” appears), protective conductor resistance is too high or the protective
conductor is interrupted. Eliminate the interruption (in the cable or at the NETZ-VDE switch).
11.4 Testing the Integrated RCD (residual current circuit breaker)
Test on a regular basis. The integrated RCD (residual current circuit breaker) can be tested
by pressing the test key. Breaking current value and time can be measured with test
instruments for DIN VDE 0413, part 6.
11.5 Fuse Replacement
All fuses can be accessed from outside. Only fuses with the breaking characteristics and
rated current values specified on the front panel may be used.
11.6 Return and Environmentally Sound Disposal
The
METRA
TESTER 5+
⏐
3P
is a category 9 product (monitoring and control instrument) in
accordance with ElektroG (German electrical and electronic device law). This device is not
subject to the RoHS directive.
In accordance with WEEE 2002/96/EG and ElektroG, we identify our electrical
and electronic devices (as of Aug. 2005) with the symbol in accordance with
DIN EN 50419 which is shown at the right. Devices identified with this symbol
may not be disposed of with the trash.
Please contact our service department regarding the return of old devices (see
address below).