CONDENSED INSTRUCTIONS:

 

 

 

Belt: 

Lay the belt around your waist and adjust it. 

 

« 

Ω 

»

 terminal

:

 

Connect the measurement coil cable to 

this terminal. This in turn must be connected to a 

telescoping rod or a test probe. 

 

4 mm banana plug: 

Connect an alligator clip or a PE 

adapter for electrical outlets to this plug. 

 

« 

ϕ 

»

 key: 

Press and hold this key until you hear a double 

acoustic signal. 

 

Alligator clip or PE adapter for electrical outlets: 

Connect this (these) to the selected equipotential bonding 

reference measurement point (e.g. the ground pin of an 

electrical outlet). 

 

Test probe or telescoping rod: 

Connect this (these) 

to the protective conductor. 

 

Display: 

Read resistance (in ohms) which was measured 

between the alligator clip (or the PE adapter for electrical 

outlets) and the test probe (or the telescoping rod).

 

 

TECHNICAL DATA:

 

 

 

Protection is impaired if the instructions are not complied with.

 

 

 

Device protection: 

400 V~ between the two device terminals or between one 

terminal and the cable plug. Fuse: 6.3 mm x 32 mm, HPC 30 kA, 500 mA~, 
690 V~. Refer to section K regarding fuse replacement.

 

 

 

User protection (with regard to grounding):

 

 

600 V~ CAT III / 300 V~ CAT IV, reinforced insulation, class 2, pollution degree 

2, per EN / IEC61010-1:2010, and 300 V~ CAT III / 300 V~ CAT IV, reinforced 
insulation, class 2, pollution degree 3, per EN / IEC61010-1:2010.

 

 

IP2X per EN / IEC60529:2001. 

Where protection of the METRALINE RLO-CHECK and accessory components 
is concerned, connection of the METRALINE RLO-CHECK to accessory 

components offers the weakest protection. 
 

Ambient conditions: 

Pollution degree 2 (conventional environment) or 3 (see 

above), temperature range for operation and storage: -20 C to +55 C, maximum 
relative atmospheric humidity: 80% for temperatures up to 31 C, linear 
reduction of relative atmospheric humidity down to 50% at 40 C. The test 

instrument may not be immersed. Protect against liquids, rain and in general 
against influences due to weather. The device may not be used in potentially 

explosive atmospheres or moist environments. 
 

Max. amperage within the cable (when used only as a cable retractor, 

see I): 

0.5 A (fuse protected).

 

 

Warning regarding interference voltage:

 

 

If interference voltage is applied to the protective conductors, resistance 
measurement is disabled. An intermittent acoustic signal is generated, the color 

of the display is changed to red and the interference voltage value is displayed. If 
a minus sign (-) appears to the left of this value, the interference voltage is direct 
voltage and otherwise alternating voltage. The METRALINE RLO-CHECK 

displays the TRMS value (superimposed, sinusoidal alternating voltage). Input 
impedance: 300 k

Ω. 

Warning range: 8 V~ / GS to 400 V~ / GS.

 

 

 
 
 

 

Maximum compensation value:

 4 

Ω

.  

 

Power supply: 

4 AA alkaline batteries (rechargeable AA batteries can be used 

but they reduce the duration of use).

 

 

 

Duration of use: 

Approx. 1.7 days under normal conditions of use at 20 C 

with new batteries. Approx. 5 hours and 50 minutes for continuous 
measurement of volume resistance of an earthing conductor of less than 1 

Ω

 

with activated acoustic signal and new batteries. The instrument is switched off 

automatically after five minutes of nonuse. 
 

Method for measuring volume resistance of the earthing conductor: 

Current between 200 and 210 mA, voltage between 4 and 6 V, automatic 
polarity reversal, in accordance with EN / IEC61557-4:2007. 

 

Functional safety of volume resistance per EN61557-4: 

±0.07 

Ω 

from 0.00 

Ω 

to 0.50 

Ω, ±10 % ±0.02 Ω 

from 0.50 

Ω 

to 2.00 

Ω  

and ±8 % at above 2.00 

Ω.  

 
In order to assure that measuring accuracy is not impaired in the following cases:

 

 

- Interchanging of accessories, 

- Relocation to another environment (significant change in temperature or 

relative atmospheric humidity), 

- Swapping of the cable plug connections (connection to the cable plug instead of 

the « AUX » terminal and vice versa). 

 

Balance the utilized accessories before performing the respective measurement 
(see sections E and G), and replace the batteries if the battery symbol blinks. 

 

Volume resistance measuring range:

 

 0.00 

Ω 

to 20.0 

Ω. 

 

Display resolution for volume resistance:

 

 

0.00 

Ω 

to 9.99 

Ω 

and 10.0 

Ω 

to 20.0 

Ω.  

 

3-447-019-03 

1/6.18 

OPERATING INSTRUCTIONS 2/2 

METRALINE RLO-CHECK 

Digital, low resistance measuring instrument 
per EN 61557-4 / VDE 0413-4 with cable reel 

and 30 m cable for measuring the resistance 
of protective conductors and equipotential 
bonding conductors 

Summary of Contents for METRALINE RLO-CHECK

Page 1: ...ed by a continuous acoustic signal indicate that balancing has been successfully completed the displayed value may fluctuate due to measuring tolerance for example between 0 00 and 0 07 The METRALINE RLO CHECK is ready for measurement The results of this balancing procedure are retained until balancing is conducted again even if the battery is replaced or the METRALINE RLO CHECK is turned off Red ...

Page 2: ...s Approx 5 hours and 50 minutes for continuous measurement of volume resistance of an earthing conductor of less than 1 Ω with activated acoustic signal and new batteries The instrument is switched off automatically after five minutes of nonuse Conformity with standards EN IEC61010 1 2010 EN IEC61557 4 2007 Conformity with European directives 2011 65 EC RoHS directive 2006 95 EC low voltage direct...

Page 3: ...r see I 0 5 A fuse protected Warning regarding interference voltage If interference voltage is applied to the protective conductors resistance measurement is disabled An intermittent acoustic signal is generated the color of the display is changed to red and the interference voltage value is displayed If a minus sign appears to the left of this value the interference voltage is direct voltage and ...

Page 4: ...for at least 2 seconds in order to switch the device off Remove the 2 Ω calibration jumper M2 Replace the battery compartment lid J2 and retighten the screw J1 Press and hold the ϕ key until you hear a double acoustic signal in order to switch the device back on again Balance resistance to your accessories in accordance with section E or G The METRALINE RLO CHECK is ready for measurement Recalibra...

Reviews: