Gossen MetraWatt METRALINE RCD CHECK Operating Instructions Manual Download Page 5

GMC-I Messtechnik GmbH

5

Tripping During Touch Voltage Measurement

If the display shown in figure 3.7 appears after pressing the 

START

 

key, the RCD has been tripped during touch voltage measure-
ment. This can be caused by an incorrectly adjusted nominal cur-
rent value 

I

N

, by a defective RCD or by previously existing leak-

age current via the RCD. This status is displayed until one of the 
function selection keys is pressed, or the device is once again 
connected to mains power.

Figure 3.7:

Example of Display for Tripping During Touch 
Voltage Measurement

Low Battery Voltage

If battery voltage is too low (only the red field is lit up in the battery 
symbol), measurement cannot be triggered with the 

START

 key. If 

the key is nevertheless pressed, the depleted battery symbol is 
displayed for about 1 second. The instrument is then returned to 
its previous status. Replace the batteries with new ones or 
recharge them as described in section 5.1.

Overheating

If several measurements with high 

I

N

 values are performed one 

after the other, the red field indicates internal warming of the 
instrument. As temperature increases, the red field is gradually 
filled in and becomes wider. An inversely displayed 

STOP

 symbol 

appears when the maximum permissible temperature is 
exceeded. If the 

START

 key is pressed, the overheating symbol 

appears for about 1 second and further measurements are dis-
abled. The instrument is then returned to its previous status which 
prevailed before the 

START

 key was pressed. Allow the test instru-

ment to cool down, which is indicated by the shrinking tempera-
ture display. 

Performing the Measurement

Securely contact the device under test with the test probes before 
triggering the measurement by pressing the 

START

 key. After-

wards, check to see whether or not the displayed line voltage 
value is stable. Do not remove the test probes from the device 
under test during measurement. Measurement results might oth-
erwise be distorted.

Influences on Measurement Results

Current flowing via PE has an unfavorable influence on mea-
surement results, or makes measurement entirely impossible, 
due to, for example, consuming devices or capacitance 
between L and PE. Consuming devices of this sort should 
therefore be removed prior to measurement.

Measurement results can be influenced by several factors: 
– Leakage current in the measured electrical circuit 
– Interference voltage occurring in an earthing system 
– When the earthing system is influenced by the potential 

of another earthing system 

– Fluctuating line voltage

If the symbol shown at the right appears together with 
the measurement results, the measurement results for 
the respective function lie within the specified limit val-
ues. 

3.3

Adjustable Parameters

The 

limit value for touch voltage

 

Ucmax

 can be set to either 50 V 

or 25 V. Adjust with the 

xI

N

 key when the touch voltage mea-

surement function (Uc) has been activated with the 

FUNC

 key.

The 

nominal value for residual current

 can be set to one of the 

following values with the 

I

key: 10 mA, 30 mA, 100 mA, 

300 mA or 500 mA.

The 

residual current multiplier

 can be set to a value of ½, 1, 2 or 

5 for 10, 30 and 100 mA, or to ½ or 1 for 300 and 500 mA 
with the 

xI

N

 key, after the 

TIME

 function has been activated 

with the 

FUNC

 key.

RCD type:

 Residual current waveform and polarity are selected 

with the 

TYPE

 key.

 

Refer to the technical data in section 4 for further details.

Figure 3.5:

Touch Voltage 
> 50 V

Figure 3.6:

Touch Voltage 
> 25 V

Figure 3.8:

Battery Voltage 
Too Low

Figure 3.9:

Battery Status 
After Pressing the 
START Key

Bild 3.10

STOP:

 Indication 

of Overheating

Figure 3.11: Indication of Over-

heating After 
Pressing the 
START Key

Display 
Symbol

RCD Type 
(residual current waveform)

Initial Polarity of the Re-
sidual Current

AC (sinusoidal)

Positive half-wave

Negative half-wave

A (pulsed)

Positive half-wave

Negative half-wave

Display 
Symbol

RCD Type

RCD Type

AC or A

Standard

Selective

TEST

S

Summary of Contents for METRALINE RCD CHECK

Page 1: ...Operating Instructions METRALINE RCDCHECK RCD Test Instrument 3 349 694 03 3 6 13 ...

Page 2: ... electrical voltage Warning concerning a source of danger attention observe documentation EC mark of conformity This instrument fulfills all require ments of applicable European directives Before using the instrument it must be assured that it is safe Do not use if Visible damage is apparent The battery compartment lid is missing The device has been stored under unfavorable conditions for a length...

Page 3: ...is enclosed in a compact housing with a pat ented means of retaining the test probes The high contrast four color OLED display assures excellent legi bility When performing measurements under unfavorable light conditions measuring point illumination can be switched on white LED at the front The RCD test instrument is capable of the following measure ments RCD disconnection time RCD breaking curren...

Page 4: ... the test probes The instrument is shutdown automatically if it has remained inactive for about 1 minute i e if no keys have been pressed and no voltage has been applied to the test probes dur ing this time 3 2 Instructions and Principles with Validity for All Measurements Setting Parameters The desired functions or parameters are selected with the FUNC I N TYPE and xI N keys see section 3 3 Measu...

Page 5: ...bes from the device under test during measurement Measurement results might oth erwise be distorted Influences on Measurement Results Current flowing via PE has an unfavorable influence on mea surement results or makes measurement entirely impossible due to for example consuming devices or capacitance between L and PE Consuming devices of this sort should therefore be removed prior to measurement ...

Page 6: ...EN 61009 and IEC 60364 4 41 standards An RCD may not trip at I N Select the TIME function by pressing the FUNC key The nominal value for residual current can be set with the I N key RCD type is selected with the TYPE key Select a multiplier with the help of the xI N key Figure 3 15 Example Settings for Time to Trip Measurement Connect the instrument to L and PE downstream from the RCD as shown in ...

Page 7: ...e to earth must be 190 V at 45 to 65 Hz dis played results are otherwise not reliable Querying the Firmware Version Before querying the firmware version disconnect both test probes from the device under test measuring circuit and switch the test instrument off Press and hold the FUNC key while switching the test instru ment on by pressing and holding the START key The firmware version appears at t...

Page 8: ...humidity 40 to 60 Device position any Ambient Conditions Operating Conditions Operating temperature 0 to 40 C Relative Humidity max 85 no condensation allowed Device position any Storage Conditions Temperature 10 to 70 C Relative Humidity max 90 at 10 to 40 C max 80 at 40 to 70 C Device position any Power Supply Batteries 4 ea AAA LR03 1 5 V alkaline or 1 2 V NIMH with at least 750 mAh Number of m...

Page 9: ...harger is con nected to the mains and to the charging socket see figure 2 1 Charging takes approximately 5 hours and 30 minutes integrated safety timer if the batteries have been fully depleted Safety Precautions Do not attempt to recharge alkaline batteries they may leak explode etc The test instrument may be severely damaged or destroyed as a result After initially charging new batteries and aft...

Page 10: ...ned deviations are used to correct display values during later use of the instrument We would be happy to perform DAkkS or factory calibration for you at our calibration laboratory Further information is available at our website www gossenmetrawatt com Company DAkkS Calibration Center or FAQs General Calibration Questions and Answers Recalibration of your instrument at regular intervals is essenti...

Reviews: