Operating Instructions

METRA

CLIP

62

Clip-On Ammeter

 3-349-391-15

2/3.10

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for METRACLIP 62

Page 1: ...Operating Instructions METRACLIP 62 Clip On Ammeter 3 349 391 15 2 3 10 www com information itm com 1 800 561 8187 ...

Page 2: ...2 GMC I Messtechnik GmbH Operating Overview 1 2 15 4 5 6 7 8 9 3 14 13 12 11 10 U 300 V I max 10 A www com information itm com 1 800 561 8187 ...

Page 3: ...witch on instrument LCD test switch off instrument 9 LCD panel 10 Eyelet for carrying strap 11 ACA key activate AC measurement 12 DCA key activate DC measurement 13 RANGE DOWN key reduce measuring range 14 DATA HOLD key save measured value by freezing display 15 ZERO ADJ key DCA zero balancing activate before each DC measurement ACA define reference value Standard Equipment 1 Clip on ammeter 1 Imm...

Page 4: ...uring Line Current 12 3 Characteristic Values 13 4 Maintenance 16 4 1 Battery Replacement 16 4 2 Housing 17 1 Safety Precautions The METRACLIP 62 clip on ammeter has been manufactured and tested in accordance with safety regulations IEC 61010 1 EN 61010 1 VDE 0411 1 and IEC 61010 2 032 EN 61010 2 032 VDE 0411 2 032 If used for its intended purpose the safety of the user and of the device is assure...

Page 5: ...a source of danger Attention observe documentation Continuous doubled or reinforced insulation Indicates EC conformity Applying the clip to dangerous active conductors is permitted The clip on ammeter may not be used If the battery compartment lid has been removed If visible damage is apparent With damaged connector cables If it no longer functions flawlessly After lengthy periods of storage under...

Page 6: ...ay not be extended beyond the safety collar during measurement Avoid excessive mechanical stress such as impact and vibration as well as high temperatures moisture and strong magnetic fields The battery compartment lid must be secured with the appropriate Phillips head screw 1 1 Special Features of the Hall Effect Sensor Eliminating the influence of magnetization The clip on ammeter METRACLIP 62 u...

Page 7: ...t sensor Remedy in the Event of Exceeding the Measuring Range If the maximum measurable current has been exceeded by a large degree zero balancing can no longer be effected due to the satura tion of the current clip Remedy Turn the measuring instrument on with the POWER key Adjust the smallest measuring range with the RANGE DOWN key 200 mA AC Close the clip around a conductor with 10 A AC and redu...

Page 8: ...reading of display values the DATA HOLD key can be activated in order to freeze the momentary measured value at the display The DH symbol appears at the display Before starting any new measurements the DATA HOLD key must be activated once again in order to reset the data hold function Display of Extreme Value Function MAX HOLD MIN HOLD In order to indicate the highest MAX HOLD or lowest current me...

Page 9: ... set to DC measurement It is therefore gener ally necessary to press the ZERO ADJUST key after selecting the measuring range and before starting DC measurements After pressing the ZERO ADJ key the symbol ZERO ADJ is displayed and 0 0 0 or 0 00 appears on the LCD After changing the measuring range or the current type the function must be reactivated Reference Value for AC measurements ACA If the ZE...

Page 10: ...asuring range switch to the highest of the three ranges in order to avoid possible over ranging Prior to DC measurements Perform zero balancing via the ZERO AD JUST key Close the clip around the current conduct ing cable under test as well as around the current transformer Read the measured value from the dis play After measurement has been completed Remove the clip from the cable and turn the mea...

Page 11: ...uring range switch to the highest of the three ranges in order to avoid possible over ranging Prior to DC measurements Perform zero balancing via the ZERO AD JUST key 1 Phase System Close the clip around both conducting cables 3 Phase System Close the clip around all 3 conducting cables Read the measured value from the dis play After measurement has been completed Remove the clip from the cable an...

Page 12: ...et the measuring range switch to the highest of the three ranges in order to avoid possible over ranging Prior to DC measurements Perform zero balancing via the ZERO AD JUST key Close the clip around a single current con ducting cable only Read the measured value from the dis play After measurement has been completed Remove the clip from the cable and turn the measuring instrument off by pressing ...

Page 13: ... 23 C 5 C Relative humidity max 80 Meas Range Input Reso lution Accuracy at 18 28 C max 80 relative humidity A AC TRMS with automatic zero balancing 200 mA 0 199 9 mA 100 A 1 0 rdg 5 digits 1 2000 mA 0 1999 mA 1 mA 1 0 rdg 5 digits 1 10 A 0 9 99 A 0 01 A 1 0 rdg 10 digits 1 A DC with automatic zero balancing 200 mA 0 199 9 mA 100 A 1 0 rdg 3 digits 2 2000 mA 0 1999 mA 1 mA 1 0 rdg 3 digits 2 10 A ...

Page 14: ...ervice life with alkaline manga nese batteries approx 120 hours Battery test Symbol B is dis played automati cally if battery voltage is too low Automatic shutdown The device turns itself off automati cally 10 minutes after it has been switched on Electrical Safety Dielectric strength 2 3 kV AC max 1 min between the clip jaws and the exposed clip hous ing Measuring category 300 V CAT III 10 A www ...

Page 15: ...ditions Operating temperatures 0 C 50 C 80 humidity no condensation allowed Storage temperatures 20 C 60 C 75 humidity no condensation allowed without batteries Mechanical Design Clip opening max dia 18 mm Dimensions W x H x D 76 mm x 194 mm x 30 mm Weight approx 350 g with batteries www com information itm com 1 800 561 8187 ...

Page 16: ...ention Remove the clip from the measuring circuit first Only then may the battery compart ment lid be removed in order to replace the batteries Set the on off switch POWER to the OFF position Remove the screw size 0 which is used to secure the battery compartment lid Pull the battery compartment lid off in the direction of the arrow symbol OPEN Carefully pry the batteries out of the com partment w...

Page 17: ...n the recesses provided Reinsert and tighten the screw Note The measuring instrument does not include any replaceable fuses 4 2 Housing No special maintenance is required for the housing Keep outside surfaces clean Use a slightly dampened cloth for cleaning Avoid the use of cleansers abrasives and solvents www com information itm com 1 800 561 8187 ...