GOSSEN-METRAWATT GMBH
7
4
Technical Data
Light Detector
silicon photodiode with V(
λ
) filter per DIN 5032/7, class C
Error Limits
per DIN 5032, part 7, class C
Most Important Error Limits
Measuring Rate
approx. 2.5 measurements per
second
LCD Field
50 mm x 25 mm
Display
7 segment
Char. Height
13 mm
Digits
3½ places
Overload Display
“OL.” is displayed
Measured Quantity
Measuring Range
Resolution
in lx
Resolution
in fc
in Lux (lx)
in footcandles (fc)
Illumination
I
0.1
…
199,9
0.01
…
19,99
0.1
0.01
I I
1
…
1
999
0.1
…
199,9
1
0.1
I I I
10
…
19
990
1
…
1
999
10
1
I V
100
…
199 900
10
…
19
990
100
10
in Candela/m²
in cd/m²
Luminance
(with luminance at-
tachment for cd/m
2
)
I
1
…
1
999
1
I I
10
…
19
990
10
I I I
100
…
199
900
100
I V
1 000
…
1 999 000
1000
Characteristics
Allowable Error
Limits per
DIN 5032 Class C
Actual Error,
MAVOLUX 5032C
V(
λ
) Matching
9%
7.5%
True Cosine Evaluation
6%
2%
Linearity
5%
1.5%
Adjustment Error
2%
1%
Total Error
20%
15%