Applications

The MAVO-SPOT  attached to the  measuring  instrument   MAVO-MONITOR provides high precision measurements
with a measuring angle of 1 degree.
The combined set permits  measuring the luminance, non- contact, at  distances  from 1 m to 

.The MAVO-SPOT is

equipped with SLR (single  lens reflex) optics having  a viewing area of  15°  and a clearly defined 1 degree measuring
circle in the centre.
Two close-up lenses available as  optional accessories permit reducing the measuring distance down to 34 cm. The light
sensor is color corrected, i.e. its spectral  responsitivity is matched to the spectral photopic vision of the human eye (V 

λ

),

acc. to DIN 5032, Part 7, Class B.
The combined set is especially suited  for measuring:

 

Monitors,  while taking into account the ambient light, e.g.  inspection and constancy tests  at  medical imaging

systems, acc. to  DIN 6868-57 and IEC 61223-2-5

(QS-RL dt. Nov. 20, 2003)

 

Lighting of streets, tunnels,  airports

 

Lighting of  sports areas

 

Light contrast at work  stations ( Protection Laws  and Regulations)

 

Lighting in museums

 

Projection screens (Checking  the uniformity of illumination)

The Functioning of the MAVO-SPOT

2.1 

Inserting the Battery

The battery compartment is located at the front of the grip of the unit.
Slide the cover of the compartment downward, insert the battery and attach the connector clip to the new battery. Then
insert battery in the compartment.

Attention: 

Only use new batteries or rechargeable accus according to IEC 6 LF 22.

2.2

Battery Control

When the instrument is switched on the green LED on the right of the 

ON/OFF

 switch

lights up. When the battery is exhausted the red LED at the left of the 

ON/OFF

 switch

will light up. You should then replace the battery immediately.
A new alkaline-manganese battery will last for approx. 60 hours of
continuous measurement operation.

Using the MAVO-SPOT

3.1 

Prepairing the measurement

Screw the protective filter, supplied with the instrument, or one of the close-up lenses on the lens of the MAVO-SPOT

.

For For getting correct measurement values, either protective filter or one of the close-up lenses must be placed on the
lens, but please note: one item only – either the filter or one of the lenses. Pull the stray light shield out. Use the attached
cable to connect the MAVO-SPOT with the MAVO-MONITOR. Switch both units on and select a suitable measuring
range.

3.2

Measuring

View through the eye-piece of the SLR viewfinder of the MAVO-SPOTand direct the measuring circle, you see in the
viewfinder to the area to be measured. This area should be illuminated uniformly and be as large as possible as
compared to the measuring circle. Press the key 

HOLD

 

(Display Hold)

 at

The MAVO-MONITOR and read the value measured.

Accessories

4.1 

Standard equipment

 

Battery

 

Instruction Manual

 

Protective filter

Summary of Contents for MAVO-SPOT

Page 1: ...2 3 Using the MAVO SPOT 2 3 1 Preparations 2 3 2 Measuring 2 4 Accessories 2 4 1 Standard equipment 2 4 2 Optional accessories 3 4 3 Calibration Certificate 3 5 Servicing and Repairs 3 6 Technical data 3 Declaration of conformity 4 Warning Do not aim the MAVO SPOT at the sun You would risk damaging your eye and also the light sensor may be destroyed 2 Lens with protective filter or close up lens a...

Page 2: ...ownward insert the battery and attach the connector clip to the new battery Then insert battery in the compartment Attention Only use new batteries or rechargeable accus according to IEC 6 LF 22 2 2 Battery Control When the instrument is switched on the green LED on the right of the ON OFF switch lights up When the battery is exhausted the red LED at the left of the ON OFF switch will light up You...

Page 3: ... Braunschweig National Standard Institute of Germany Depending on how the instrument is being used we recommend a recalibration interval between 12 and 18 months For this purpose please contact our Calibration Service Department telephone 49 911 8602 172 5 Servicing and Repairs No special maintenance is required if the MAVO SPOT is handled correctly Keep the outside surfaces clean Use a slightly d...

Page 4: ...________________________ Ort Datum Place date Vorsitzender der Geschäftsführung Diese Erklärung bescheinigt die Übereinstimmung mit den genannten Richtlinien This declaration certifies compliance with the above mentioned beinhaltet jedoch keine Zusicherung von Eigenschaften Die Sicherheitshinweise directives but does not include a property assurance The safety der mitgelieferten Produktdokumentati...

Reviews: