11
Gossen Metrawatt GmbH
METRACABLE TDR PRO
1. Press the
ON/OFF
key at the instrument.
The instrument is switched on and the TDR trace is displayed.
2. Open the main menu by pressing and holding the
ESC
key.
The main menu appears.
3. Navigate to the
Settings
submenu with the help of the
and
keys.
4. Acknowledge by pressing the
SEL
key.
The
Settings
menu is displayed.
5. Navigate to the desired setting with the help of the
and
scroll keys.
6. Acknowledge by pressing the
SEL
key.
The respective submenu is displayed.
7. Select the desired option or enter the desired value.
8. Acknowledge by pressing the
SEL
key.
The setting is saved.
9. Return to the measurement function by pressing the
ESC
key.
The instrument is ready to perform measurements.
4.6
Changing Measurement Settings
The following settings are required for measurements:
•
AC or DC input
•
Pulse length
•
Cable impedance
Detailed information concerning the settings and their values
These settings should be selected before performing a measurement. However, if necessary
they can also be changed during a measurement. The measurement is restarted automati-
cally after each change.
1. Press the
ON/OFF
key at the instrument.
The instrument is switched on and the TDR trace is displayed.
2. Open the TDR menu by pressing and holding the
SEL
key.
The TDR menu appears.
3. Navigate to the desired setting with the help of the
and
scroll keys.
4. Acknowledge by pressing the
SEL
key.
The respective submenu is displayed.
5. Select the desired option or enter the desired value.
6. Acknowledge by pressing the
SEL
key.
The setting is saved.
7. Return to the measurement function by pressing the
ESC
key.
The measurement is updated automatically based on the changed settings.