Report Functions
Depending on the selected operating mode,
the measured values and the result of each
test can be transmitted to a PC via USB port
for onward processing or stored to the inter-
nal device memory. The transmission or
storage process of measured values is indi-
cated by all LEDs lighting briefly up from
right to left.
Selection of Operating Mode
The corresponding PC analysis software
programs have to be installed and started
for the three different operating modes in
order to receive data, see table above.
Press key „S“ (6) when switching on the
instrument to set the requested operating
mode. The following operating modes are
shown in subsequent order:
•
„Transmission mode“, siehe chapter 6.1
•
„Permanent transmission“, see chapter 6.2
•
„Memory mode“, see chapter 6.3
The operating mode which is indicated when
releasing key „S“ (6) is selected. This setting
remains active even after switching off the
instrument.
6.1 Transmission Mode (Transmission only
while Program MINITEST is Active)
To transmit the current measured value, key
„S“ must be pressed in operating mode
„Transmission mode“ .
6.2 Permanent Transmission (Transmission
only while Program MINITEST is Active)
If operating mode „Permanent transmission“
is activated, each measured value is trans-
mitted via the USB port.
6.3 Memory Mode (Read-Out only with
Program MINI
TEST
⏐
M
ASTER
)
In the test instruments
MINI
TEST
⏐
M
ASTER
or
MINI
TEST 3P
⏐
M
ASTER
a maximum of 2,047
DUTs with 10 measured values per DUT can
be stored. If an attempt is made to store
more than 10 measured values for one DUT,
the additional values are ignored and the red
LED lights up briefly.
6.3.1 Activate Memory Menu
After selecting memory mode (see above) you
have to set the rotary switch to position Gerä-
teselbsttest/Speichermenü
(Self-test/Memory
Menu)
. Activate the memory menu by pressing
key „S“ (6). The following is shown on the LCD:
** MEMORY MENU ** SELECTION -> SWITCH
.
You may now select the requested memory
function via the rotary switch.
6.3.2 Selecting Memory Function
➭
Key NUM – > Menu DUT NUM –
The number of the current memory location
is reduced by one each time key „S“ (6) is
pressed. The lowest possible value is „1“.
The letter „D“ in front of the number signifies
that data are already available for this mem-
ory location.
➭
Key NUM + > Menu DUT NUM+
The number of the current memory location
is increased by one each time key „S“ (6) is
pressed. The highest possible value is
„2,047“, i.e. a maximum of 2,047 DUTs can
be tested. The letter „D“ in front of the num-
ber signifies that data are already available
for this memory location.
➭
Key DATA > Menu MEASURED VALUE NO.
In this switch position all measured values of
a DUT are shown in a consecutive, numeri-
cal sequence in the order of their recording.
Operating mode selected
at the test instrument
MIN
I
TEST 3
P
⏐
M
ASTE
R
MIN
I
TEST
⏐
M
ASTE
R
MIN
I
TEST
⏐
P
RO
PC
analysis
software
depend-
ing on the
operating
mode
Transmission mode
(Transmission of individual
measured values)
✘
✘
✘
MINITEST
Permanent transmission
(Continuous transmission of
measured values)
✘
✘
✘
MINITEST
Memory mode
(Internal measured value
memory)
✘
✘
—
MiniTEST
Master