1 Installation and startup
1.1 Scope of delivery
(see figure 1)
The continuous-line recorder LINAX 4000M comes with:
– 1 copy of operating instructions
– 2 fasteners
Be
– 1 fiber pen insert
Fe
per measuring channel
– 1 print insert
De
(Option)
– 1 pack of fanfold chart
Fp
or 1 roll chart
Sr
– 24 jumpers for universal version
– Depending upon the order, the respective number of screw-
plug terminals
Sk
, Sub-D plug, 9-pin, and reading ruler(s).
Figure 1
Scope of delivery of the LINAX 4000M
1.2 Selecting the mounting site
Figure 2
Dimensional drawing LINAX 4000M (dimensions in mm)
Position of use
Inclination to the side –30°... 0 ... + 30°
Inclination to the rear 20°
Inclination to the front 20°
Ambient temp.
0 ... 50
°
C
Relative humidity
≤
75 % annual average,
max. 85 %.
Prevent dewing!
1.3 Installation
(see figure 2 and figure 3)
Installation in switchboards
1. Insert the recorder into the switchboard from the front.
2. At the sides of the case, slide the fasteners
Be
into the guide
grooves (see figure 3).
Note
The fasteners
Be
are suited for side-by-side mounting in hori-
zontal or vertical direction.
3. After aligning, equally tighten the fasteners
Be
.
Figure 3
Mounting the fasteners
Be
Fastener
De
Print insert (Option)
Fe
Fiber pen inserts
Fp
Fanfold chart
Sk
Screw-plug terminals
Sr
Chart roll
Fe
Sk
Be
Sr
Fp
De
Panel cutout