GPT-9600 Series Specifications
AC Withstanding Voltage
[for all models]
Output-Voltage Range
0.10kV~ 5.00kV
Output-Voltage Resolution
10V
Maximum Rated Load
100 VA (5kV/20mA)
Maximum Rated Current
20mA (0.5kV
<
V 5kV)
≦
, 5mA (0.1kV V
≦ ≦
0.5kV)
Output-Voltage Waveform
Sine wave
Output-Voltage Frequency
50 Hz / 60 Hz selectable
Voltage Regulation
±
(1.5% + 2 counts) [full load ->no load]
Voltmeter Accuracy
±
(1.5% of r 2 counts)
Current Measurement Range [1]
0.01mA~20.0mA
Current Best Resolution
0.01mA(0.01mA~9.99mA)
0.1mA(10.0mA~20.0mA)
Current Measurement Accuracy
± (2.0% of rdg + 5 counts) when HI SET<1.00mA
± (2.0% of rdg + 3 counts) when HI SET
≥
1.00mA
Window Comparator Method
Yes
ARC Detect
Yes
RAMP (Ramp-Up Time)
0.1s Fixed
TIMER (Test Time)
OFF, 1s~180s
GND
ON
[ 1 ] Output Limitation in AC Withstanding Voltage Testing as below.
Upper Current
Pause
Output Time
15mA
≦
I
≦
20mA
At least as long as the output time
Approx. 180 seconds
AC
0.01mA
≦
I<15mA
Not necessary
Continuous output possible
DC Withstanding Voltage
[for GPT-9602/9603]
Output-Voltage Range
0.10kV~ 6.00kV
Output-Voltage Resolution
10V
Maximum Rated Load
25W (5kV/5mA)
Maximum Rated Current
6mA (0.5kV
<
V 6kV)
≦
, 2mA (0.1kV V 0.5kV)
≦ ≦
Voltage Regulation
±
(1.5% + 2 counts) [full load ->no load]
Voltmeter Accuracy
±
(1.5% of r 2 counts)
Current Measurement Range [1]
0.01mA~6.00mA
Current Best Resolution
0.01mA(0.01mA~6.00mA)
Current Measurement Accuracy
± (2.0% of rdg + 5 counts) when HI SET<1.00mA
± (2.0% of rdg + 3 counts) when HI SET
≥
1.00mA
Window Comparator Method
Yes
ARC Detect
Yes
RAMP (Ramp-Up Time)
0.1s Fixed
TIMER (Test Time)
OFF, 1s~180s
GND
ON
[ 1 ] Output Limitation in DC Withstanding Voltage Testing as below.
Upper Current
Pause
Output Time
DC
0.01mA
≦
I
≦
6.00mA
Not necessary
Continuous output possible